Reviewing data from a data card

If a data card is installed when the HeartStart FR2+ is turned on for use during an incident, the HeartStart FR2+ automatically records detailed information on the data card. To review this information on the HeartStart FR2+ screen:

1.Make sure the training & administration pack is not installed.

2. Make sure the data card is installed. Unplug the pads connector.

3. Remove and reinstall the battery.

4. Select REVIEW INCIDENT from the menu. A new screen comes up. This screen displays:

￿ELAPSED TIME — how long the incident recorded by the FR2+ lasted,*

￿SHOCKS DELIVERED — how many shocks were delivered during the incident, and

￿FIRST SHOCKS AT — the times at which the first three shocks were delivered.

N O T E : If the data card does not contain event data, only the summary information from FR2+ internal memory will be displayed when REVIEW INCIDENT is selected.

5.To review the events that occurred during the incident, select REVIEW

EVENTS. A new screen comes up. This and following screens, accessed by selecting NEXT EVENTS, display elapsed time information for critical activities in using the FR2+. These include:

￿ POWER ON — when the FR2+ was turned on,

￿PADS ON — when the defibrillator pads were connected,

￿SHOCK ADVISED — when a shock was advised,

￿ARMED — when the FR2+ charged for shock delivery,

￿SHOCKED — when a shock was delivered,

￿SHOCK ABORTED — when a shock was aborted,

￿PAUSE FOR CPR — when a pause occurred

￿POWER OFF — when the FR2+ was turned off

*The FR2+ displays elapsed time to a maximum of 99:59 minutes. If the elapsed time of use extends beyond this figure, the minutes are represented by “??” but the seconds are displayed. However, total elapsed time will be recorded on an installed data card for later review with HeartStart Event Review data management software.