symptom
General troubleshooting
possible cause | recommended action |
Status Indicator:
FLASHING RED X
Audio signal:
CHIRPING
•The energy remaining in the battery is low.
•The FR2+ has been stored outside the recommended temperature range.
•An error has been detected as part of the
•The FR2+ has been unable to perform its daily
•Replace battery with a new M3863A FR2 standard or a charged M3848A FR2+ rechargeable battery as soon as possible.
•Remove and reinstall the battery and run a battery insertion
has been stored outside the recommended temperature range. See Appendix B for recommended range.
•Remove and reinstall the battery and perform the battery insertion
•Make sure defibrillator pads are not attached to the FR2+.
Status Indicator:
FLASHING OR SOLID RED X
Audio signal:
CHIRPING
Screen prompt (displayed for 10 seconds at the end of a BIT, before FR2+ turns off):
NOT READY FOR USE
or
SELFTEST FAILED
A test revealed a failure or error. The FR2+ performs
•Unplug the pads connector from the FR2+, if connected.
•Remove and reinstall the battery and check the results of the battery insertion
NOTE: You can stop the tests and use the FR2+ as soon as you see the Status Indicator change to the flashing black hourglass. Simply press the On/Off button to stop the test and put the FR2+ into standby mode. The FR2+ is then ready for use.