Philips M3860A, M3861A manual Receiving setup

Models: M3860A M3861A

1 102
Download 102 pages 22.88 Kb
Page 55
Image 55

P hili ps Me dical Sy stems

6-7

￿Use the infrared communications feature of the FR2+ to receive the revised setup from another FR2+. (Instructions are provided later in this chapter.)

￿Use the infrared communications feature of the FR2+ to receive the revised setup from a computer running HeartStart Event Review software. (Instructions are provided with the HeartStart Event Review software.)

C A U T I O N : Alteration of the factory default setup of the FR2+ can affect its performance and should be performed under the authorization of your Medical Director. Modifications to device operation resulting from changes to the default settings should be specifically covered in user training.

See the tables describing the various setup parameters at the beginning of this chapter and Appendix D, “Glossary of Terms,” for definitions of setup items.

Receiving setup

This method uses the infrared communications feature of the HeartStart FR2+ to receive setup directly from one HeartStart FR2+ to another (which must have the training & administration pack installed in it) or from a computer running HeartStart Event Review software. (See instructions provided with HeartStart Event Review.) To receive setup from another FR2+, follow these steps:

1.Locate the infrared communications port on each HeartStart FR2+ and

line them up with one another, so that the infrared “eye” in each one has an uninterrupted view of the “eye” in the other. (See the diagram on the inside front cover.) The two devices should be no more than 1 meter 6 apart.

2. Make sure the “sending” FR2+ has the training & administration pack installed and is ready to send. (See the M3864A training & administration pack Reference Guide for instructions.)

3.Select RECEIVE SETUP from the setup menu:

4.A new screen comes up. Until the two HeartStart FR2+ devices are

properly positioned, the screen displays READY TO RECEIVE and prompts you to check the sending FR2+.

5. Setup data are automatically transferred as soon as the infrared ports are correctly aligned.

6. If you select EXIT before the transfer is complete, the revised setup will not be received. When the transfer is complete, the screen on the

Setup and Advanced Mode Features

Page 55
Image 55
Philips M3860A, M3861A manual Receiving setup