Philips M3860A, M3861A manual Optional M3848A FR2+ rechargeable battery

Models: M3860A M3861A

1 102
Download 102 pages 22.88 Kb
Page 75
Image 75

P hili ps Me dical Sy stems

 

B-7

B

 

(Optional) M3848A FR2+ rechargeable battery

 

 

 

 

 

category

nominal specifications

 

 

 

 

Battery Type

11.3 VDC, 6.5 Ah, lithium ion. Rechargeable cell using the M3849A charger.

 

 

 

 

Capacity

When freshly charged and used at 77° F (25° C) , provides a minimum of 80

 

 

(typically 100) shocks, or 3.5 hours (typically 5 hours) of ECG display time only,

 

 

before recharging is indicated.

 

 

 

 

Status Indicators

Good battery: bar graph on display screen indicating remaining power level.

 

 

Low battery: flashing red X on the front panel of the FR2+ (When low battery

 

 

indicator appears, there is still enough energy to deliver 9 shocks plus 15 minutes

 

 

of ECG display time).

 

 

Dead battery: solid red X on the front panel of the FR2+.

 

 

 

 

Storage Temperature

32° to 109° F (0° to 43° C).

 

 

 

 

Standby Life

6 months when installed fully charged in a defibrillator labeled FR2+.

 

(after installation)

 

 

 

(Optional) M3849A charger

 

 

 

 

 

 

category

nominal specifications

 

 

 

 

Application

For use with M3848A FR2+ rechargeable battery only.

 

 

 

 

Power Requirements

100 to 240 VAC, 47 to 63 Hz, 30 Watts

 

 

 

 

Environmental

32° to 122° F (0° to 50° C).

 

Requirements

0% to 95% relative humidity (non-condensing).

 

Conformance Testing

International: EN60335-1:1994 Class 1

 

 

North America: UL 1310 Class 2

 

 

M3870A and DP2/DP6 defibrillator pads

 

 

 

 

 

 

category

nominal specifications

 

 

 

 

Pads, Cable, and Connector Disposable and self-adhesive. DP2/DP6 adult defibrillator pads have a nominal active surface area of 100 cm2 each and are provided in a sealed package with an integrated 122 cm (48 inch), typical, cable and connector. M3870A FR2 infant/child reduced-energy defibrillator pads have a nominal active surface area of 44 cm2 each and are provided in a sealed package with an integrated 122 cm (48 inch), typical, cable and connector incorporating attenuating electronics.