symboldescription
High voltage.
Refer to operating instructions.
Meets the requirements of the European medical device directives.
This product has passed relevant safety tests by CSA, a nationally recognized test lab.
IP54 | With data card tray and battery installed, meets IEC 529 |
| class IP54. |
|
|
HeartStart FR2+ display screen
symbol
HR XXX
XX
XX:XX
TEMPERATURE
SETUP
REV: XXX X.X
XXXX
description
Heart rate.
Number of shocks delivered.
Time. How much time (minutes:seconds) has passed since the HeartStart FR2+ was turned on.
Temperature. Recommended storage temperature range has been exceeded since the last battery insertion
Setup. Setup has been lost from memory; factory default setup is being used. Contact Medical Director for revised setup.
Software. The version of software used in your HeartStart FR2+.