stems

7-3

To replace a data card:

I M P O R T A N T : You must turn the FR2+ off (return it to standby mode) before you remove the data card, to ensure that no incident data are lost.

1.Press the On/Off button to turn off the FR2+. Never replace the data card unless the FR2+ is turned off.

2.Remove the loaded data card tray by grasping its handle and pulling it out of the port.

3.Remove the data card from the tray.

4.Give the data card to the appropriate person in your organization.

5.Because it helps seal the FR2+ against moisture, the data card tray should always be reinserted into the port of the FR2+. Either load a new data card into the tray and insert it, or insert the empty data card tray into the port.

Reviewing Incident Data

Reviewing data from internal memory

Summary information from the last incident that is stored in the internal memory of the HeartStart FR2+ can be displayed on its screen for review. To review this information:

1.Remove the data card if one is installed and unplug the pads connector.

2. Remove and reinstall the battery. (Make sure you are using the gray M3863A FR2 standard battery or the blue M3848A rechargeable battery, not the yellow training & administration pack.)

3. Select REVIEW INCIDENT from the menu. A new screen comes up.

4. Observe and record, if desired, the summary information displayed on the screen:

dical Sy

￿ how long the incident recorded by the FR2+ lasted, and

￿ how many shocks were delivered during the incident.

7

P hili ps Me

This information stays in the FR2+’s memory and can be displayed for review until the next time the FR2+ is used. At that time, the data from the new incident will be displayed.

Data Management and Review

Page 63
Image 63