(Optional) M3873A/M3874A FR2+ ECG assessment

 

 

 

module

 

 

 

 

 

category

nominal specifications

 

 

 

 

Application

For use with the FR2+ M3860A with ECG display enabled and running version

 

 

1.5 software or higher (denoted by FR2+ on the front panel or rear label).

 

 

 

 

 

Length and Weight

100 inches (182 cm); ≤ 1 lb. (2.2 kg).

 

 

 

 

 

 

Operating Temperature and

32° to 122° F (0° to 50° C);

 

 

Humidity

0% to 95% relative humidity (non-condensing).

 

 

 

 

 

Storage Temperature and

32° to 109° F (0° to 43° C).

 

 

Humidity

0% to 75% relative humidity (non-condensing).

 

 

 

 

 

Patient Lead Wire

M3873A (AAMI):

M3874A (IEC):

 

Designation

positive lead − red

positive lead − green

 

 

negative lead − white

negative lead − red

 

 

reference lead − black.

reference lead − yellow

 

 

 

 

 

Typical (Lead II) Connection

Lead II vectors:

 

 

 

positive − left leg

 

 

 

negative − right arm

 

 

 

reference − left arm.

 

 

 

Other limb vectors can be obtained by different electrode positions.

 

 

 

 

Battery Type

3 V, 1 Ah, poly-carbonmonofluoride lithium (LiCFx). Non-replaceable disposable

 

 

primary cell.

 

 

 

 

 

 

Service Life

Typically, 5 years.

 

 

 

 

 

Performance with FR2+

Meets environmental specifications cited for FR2+ Defibrillator on page B1

 

Defibrillator

through B2.

 

 

 

 

 

 

P hili ps Me dical Sy stems

Technical Specifications