P hili ps Me dical Sy stems

 

 

 

6-5

 

 

 

 

 

 

 

 

 

 

 

 

parameter

settings

default

description

 

 

 

 

 

 

 

 

 

 

 

 

resume key

ON, OFF

OFF

Enables (ON) or disables (OFF) user-initiated interruption

 

 

 

of the CPR pause and return to analyzing. If either the

 

 

 

CPR Timer or the NSA Action setting is programmed to

 

 

 

1.5 minutes or longer, the Resume Key setting is

 

 

 

automatically enabled (ON).

 

 

 

If enabled, the Resume

 

 

 

 

 

 

 

 

 

 

 

Key is accessed by

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

pressing the lower Option

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

button indicated by an

 

 

 

 

 

 

 

 

 

 

 

arrow on the FR2+

 

 

 

 

 

 

 

 

 

 

 

display, as shown in the

 

 

 

 

 

 

 

 

 

 

 

sample screen:

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

advanced use

0.5, 1.0, 1.5,

0.5

Sets the interval for “Press to Analyze” prompts provided

prompt

2.0, 2.5, 3.0

 

during advanced mode operation.

interval

 

 

 

(minutes)

 

 

 

 

 

 

 

Using Setup Features

N O T E : To move around the menus displayed, use the Option buttons as follows:

• Press the LOWER Option button to move the highlight bar from one item to another on the menu.

• Press the UPPER Option button to select the highlighted item or to

scroll through the settings for that item.

6

 

The FR2+ comes with a factory default setup designed to meet the needs of

 

most users. The setup feature of the FR2+ lets you review the current setup

 

of your HeartStart FR2+ or install a revised setup if appropriate. To go to the

 

SETUP menu:

 

1.Remove and reinstall the battery to bring up the first menu on the screen.

N O T E : This screen will not be displayed if the FR2+ is connected to defibrillator pads (that are applied to the patient) when the battery is inserted, and you will not be able to access the menu items. In addition, the battery insertion selftest and periodic automatic selftests cannot run while the defibrillator pads are connected. Be sure to unplug the pads connector from the FR2+ after each use. Do not store the FR2+ with the pads connected.

Setup and Advanced Mode Features

Page 53
Image 53
Philips M3860A, M3861A manual Using Setup Features