M3860A/M3861A HEARTSTART FR2+
6-4
Philips Medical Systems
Manual override parameters

The HeartStart FR2+ provides several ways of overriding the automatic

protocol. The parameters in the following table are used to enable different

kinds of manual override.

parameter settings default description

advanced OFF,
ANALYZE,
CHARGE
OFF Enables or disables advanced mode entry for ALS or
tiered-response systems.
OFF — disables advanced mode features.
ANALYZE — enables user-initiated rhythm analysis and
disarm, and (M3860A only) automatically turns on ECG
display when advance mode is entered.
CHARGE (M3860A only) — in addition to enabling the
analyze feature, enables user-initiated charging and
disarming.
pause key OFF,
MONITOR,
ALWAYS
OFF Enables or disables user-initiated CPR pause in the
automatic protocol. The length of the pause is defined by
the CPR Timer setting. When an Advanced mode feature
(ANALYZE or CHARGE) is enabled and accessed, the
Pause key is disabled.
OFF — disables availability of user-initiated pause.
MONITOR — enables user-initiated pause only during FR2+
monitoring of patient rhythm.
ALWAYS — enables user-initiated pause any time except
when the device is already paused.
If enabled, the Pause Key
is accessed by pressing
the lower Option button
indicated by an arrow on
the FR2+ display, as
shown in the sample
screen: