6-4

Manual override parameters

The HeartStart FR2+ provides several ways of overriding the automatic protocol. The parameters in the following table are used to enable different kinds of manual override.

parameter

settings

default

 

 

 

advanced

OFF,

OFF

 

ANALYZE,

 

 

CHARGE

 

 

 

 

pause key

OFF,

OFF

 

MONITOR,

 

 

ALWAYS

 

 

 

 

description

Enables or disables advanced mode entry for ALS or tiered-response systems.

OFF — disables advanced mode features.

ANALYZE — enables user-initiated rhythm analysis and disarm, and (M3860A only) automatically turns on ECG display when advance mode is entered.

CHARGE (M3860A only) — in addition to enabling the analyze feature, enables user-initiated charging and disarming.

Enables or disables user-initiated CPR pause in the automatic protocol. The length of the pause is defined by the CPR Timer setting. When an Advanced mode feature (ANALYZE or CHARGE) is enabled and accessed, the Pause key is disabled.

OFF — disables availability of user-initiated pause.

MONITOR — enables user-initiated pause only during FR2+ monitoring of patient rhythm.

ALWAYS — enables user-initiated pause any time except when the device is already paused.

If enabled, the Pause Key is accessed by pressing the lower Option button

indicated by an arrow on the FR2+ display, as shown in the sample screen: