Performance Verification
 AWG510 & AWG520 Service Manual
c. Repeat substeps 4a through 4c, as checking to the follow limits:
Rise time 2.5 ns, maximum
Aberration ± 10 %, maximum
Flatness ± 3 %, maximum
6. Check CH1 or CH2: Repeat the Check Pulse Response for the AWG510
CH1 or the AWG510 CH2, depending on the instrument that you are
currently testing.
7. End procedure: Disconnect the oscilloscope.