Performance Verification
 AWG510 & AWG520 Service Manual
 #* %
(!  $'
##
&#  
 &!  
 ! $%'
'   
 &%
d. Set the signal generator controls:
 %& &$
#%#$ 
#"&)  *
!%&    !!
&%!&%  
2. Set the AWG500–Series Waveform Generator controls and select the
waveform file:
a. Initialize the AWG500–Series Waveform Generator controls:
Push UTILITY (front–panel)
System (bottom)Factory Reset (side)OK (side).
b. Select the file: Load the MODE.WFM as referring to the procedures on
page 4–15.
c. Set the AWG500–Series Waveform Generator clock source to external:
Push HORIZONTAL MENU (front–panel)Clock Src (side) so that
the clock reference is set to External.
3. Turn on the AWG500–Series Waveform Generator CH1 output: Push the
RUN and CH1 OUT buttons so that the LEDs above the RUN button and
CH1 output connector light.
4. Check against limits: Check that the amplitude of the sine wave displayed on
the oscilloscope is 5 vertical divisions and that the waveform of one cycle
per 2 horizontal divisions is displayed.
5. End procedure: Turn the signal generator output off and disconnect the
signal generator .