Performance Verification
AWG510 & AWG520 Service Manual
2. Set the AWG500–Series Waveform Generator controls and select the
waveform file:
a. Initialize the AWG500–Series Waveform Generator controls:
Push UTILITY (front–panel)
System (bottom)Factory Reset (side)OK (side).
b. Select the file: Load the MODE.WFM as referring to the procedures on
page 4–15.
c. Set trigger mode: Push SETUP (front–panel)Run Mode (bot-
tom)Triggered (side).
d. Set trigger interval:
Push SETUP (front–panel)Trigger (bottom)Source (side)
Internal (side).
Push the Interval side button.
Enter numeric value of 1 m: Push 1 and m (SHIFT+9) keys in this
order.
3. Turn on the AWG500–Series Waveform Generator CH1 output: Push the
RUN and CH1 OUT buttons so that the LEDs above the RUN button and
CH1 output connector light.
4. Check the trigger interval: Check that the period of the waveform displayed
on the oscilloscope is one horizontal divisions.
5. Change the trigger interval:
a. Enter numeric value of 2 ms: Push 2 and m (SHIFT+9) keys in this
order.
6. Check the trigger interval: Check that the period of the waveform displayed
on the oscilloscope is two horizontal divisions.
7. End procedure: Disconnect the oscilloscope.