Removal and Installation Procedures
AWG510 & AWG520 Service Manual 6-39
1. Assemble equipment and locate modules to be removed: Have handy a
screwdriver with a size Phillips #2 tip (Items 1 and 3). Find the modules to
be removed in the locator diagram Outer-Chassis Modules, Figure 6–3,
page 6–3.
2. Orient the waveform generator: Set the generator so its top is down on the
work surface and its rear is facing you.
3. Remove the A40 Clock Board.: Use Figure 6–18 as a guide.
a. Unplug two interconnect cables at J41, J200, and J340 on the A40 Clock
board.
b. Remove the four screws attaching the Clock board to the AWG board.
c. Disconnect the Clock board from the AWG board, then lift it up away to
complete the removal.



 
Figure 6-18: A40Clock Board removal
4. Reinstallation: Do, in reverse order, steps 3a through 3c reversing the
removal instructions of each step to reinstall the Clock board.
A40 Clock Board