Performance Verification
 AWG510 & AWG520 Service Manual
d. Set the function generator (AFG310) controls:
!    !
   !!
  
!"  #
 !   ! 
   
! ! 
2. Set the AWG500–Series Waveform Generator controls and select the
waveform file:
a. Initialize the AWG500–Series Waveform Generator controls:
Push UTILITY (front–panel)
System (bottom)Factory Reset (side)OK (side).
b. Select the file: Load the MODE.WFM as referring to the procedures on
page 4–15.
c. Set the AWG500–Series Waveform Generator clock reference to external:
Push HORIZONTAL MENU (front–panel)Clock Ref (side) so
that the clock reference is set to External.
Push the Clock side button.
Enter numeric value of 50M: Push 5, 0, M (SHIFT+7) keys in this
order.
3. Turn on the AWG500–Series Waveform Generator output: Push the RUN
button so that the LED above the RUN button lights.
4. Check the clock output frequency with 10.0 MHz reference: Check that the
frequency counter reading is 50.0 MHz "10 kHz.
5. Modify the function generator controls:
!"   #
6. Check the clock output frequency with 10.1 MHz reference: Check that the
frequency counter reading is 50.5 MHz "10 kHz.
7. End procedure: Turn the function generator output off and disconnect the
function generator and frequency counter.