Adjustment Procedures
 AWG510 & AWG520 Service Manual
Turn the general purpose knob to find the upper limit for the
waveform corrupted area. Note the value as N2.
Calculate (N1+N2)/2–28 and set the value to it. That is the proper
setting. For example, if N1=40 and N2=50, then set the value to 17.
The following step 5 is for the AWG520 only. For the AWG510, go to the step 6.
5. Adjust CH 2 DAC clock timing: (for AWG520 only)
a. Modify the AWG500 hookup: Disconnect the coaxial cable on the CH 1
output connector and connect it to the CH 2 output connector.
b. Set AWG500 controls: Press CH2 Delay.
c. Adjust timing:
Turn the general purpose knob to find the setting of output wave-
form corrupted. The value should be near 40.
Turn the general purpose knob to find the lower limit for waveform
corrupted area. Note the value as N1.
Turn the general purpose knob to find the upper limit for waveform
corrupted area. Note the value as N2.
Calculate (N1+N2)/2–28 and set the value to it. That is proper
setting.
6. Save the settings: Press Save.
7. End procedure: Disconnect the oscilloscope.