Adjustment Procedures
AWG510 & AWG520 Service Manual 5-7
Adjustment InstructionsThis procedure adjusts Y330 on the A40 Clock board to set the AWG500 clock
frequency.
Equipment
Required One frequency counter (Item 2)
One 50 coaxial cable (Item 5)
1. Connect the test equipment and set test equipment controls:
a. Connect the frequency counter: Connect the AWG500 Clock Output
connector through a BNC coaxial cable to the input A connector on the
frequency counter.
InputClock output
AWG500 Frequency counter
50 coaxial cable
Figure 5-3: Hookup for the clockfrequency adjustment
b. Set frequency counter controls:
Input A Coupling . ............. AC
Funciton .................... AFrequency
Gate Time . ................. 0.2s
Level . ..................... 0V
2. Set AWG500 controls:
a. Initialize AWG500 controls: Press UTILITY System Factory
Reset OK.
b. Load waveform:
Press SETUP Waveform/Sequence Load...
Turn the general purpose knob to display the list of waveform files
and highlight the file MODE.WFM.