Performance Verification
AWG510 & AWG520 Service Manual 4-55
Figure 4-24: Waveformoutput when the SW2 is closed
c. Degenerate the event signal: Open the SW2 of the ground closure to
degenerate the event signal.
d. Confirm the waveform on the oscilloscope: Confirm that the oscilloscope
displays back the waveform in Figure 4–22.
7. Check the EVENT IN connector pin 2 input:
a. Generate an event signal: Close the SW3 of the ground closure to
generate an event signal on the EVENT IN connector pin 2.
b. Confirm the waveform on the oscilloscope: Confirm that the oscilloscope
displays the waveform as shown in Figure 4–25.
Figure 4-25: Waveformoutput when the SW3 is closed