Performance Verification
AWG510 & AWG520 Service Manual 4-43
Sine Wave Tests
This procedure checks the sine wave characteristics of the AWG500–Series
Waveform Generator output waveforms.
Equipment
required
A 50Ăcoaxial cable, a DC block, an adapter (N male to BNC female),
and a spectrum analyzer.
Prerequisites The instrument mustmeet the prerequisites listed on page 4-13.
1. Install test hookup and set test equipment controls:
a. Hook up the spectrum analyzer: Connect the AWG500–Series Waveform
Generator CH1 output connector through the coaxial cable, adapter, and
DC Block to the input connector on the spectrum analyzer (see Figure
4–14).

AWG500Series Waveform Generator Spectrum Analyzer
50 Wcoaxial cable
Adapter
and DC
Block
Figure 4-14: Sine wave initial testhookup
b. Set spectrum analyzer controls:
Center frequency . ................ 200MHz
Full Span . ...................... 500MHz
Reference level . .................. 10dBm
RFattenuation . .................. 20dB
Videofilter . ..................... 1kHz
ResolutionBW . .................. 1MHz