Adjustment Procedures
AWG510 & AWG520 Service Manual 
2. Set the waveform generator controls:
a. Initialize AWG500 controls: Press UTILITY System Factory
Reset OK.
b. Load CH 1 waveform:
Press SETUP Waveform/Sequence Load...
Turn the general purpose knob to display the list of waveform files
and highlight the file TRI512.PAT.
Press OK to select and load the file.
c. Start waveform generation:
Press CH 1 OUTPUT button.
The following steps d and e are for the AWG520 only. For the AWG510, go
to the step 3.
d. Load CH 2 waveform: (for AWG520 only)
Press CH 2 button to select the channel.
Press SETUP Waveform/Sequence Load...
Turn the general purpose knob to display the list of waveform files
and highlight the file TRI512.PAT.
Press OK to select and load the file.
e. Start waveform generation:
Press CH 2 OUTPUT button.
3. Enter DAC adjust menu: Press UTILITY Service Tweak AWG1
OK.
4. Adjust CH 1 DAC clock timing:
a. Set AWG500 controls: Press CH1 Delay.
b. Adjust timing:
Turn the general purpose knob to find the setting at which the output
waveform corrupts. The value should be near 40.
Turn the general purpose knob to find the lower limit for the
waveform corrupted area. Note the value as N1.