Performance Verification
4-78 AWG510 & AWG520 Service Manual
Noise Output TestsThis procedure checks the AWG500–Series Waveform Generator noise output
characteristics.
Equipment
required
A 50ĂΩcoaxial cable, a DC block, an adapter (N male to BNC female),
and a spectrum analyzer.
Prerequisites The instrument mustmeet the prerequisites listed on page 4-13.
1. Install test hookup and set test equipment controls:
a. Hook up the spectrum analyzer: Connect the AWG500–Series Waveform
Generator NOISE OUT connector through the coaxial cable, adapter,
and DC Block to the input connector on the spectrum analyzer (see
Figure 4–37).
AWG500Series Waveform
Generator rear panel Spectrum Analyzer
50 Wcoaxial cable
Adapter and
DC Block
Figure 4-37: Noise output initial testhookup
b. Set spectrum analyzer controls:
Centerfrequency ................. 100MHz
Full span . ...................... 500MHz
Vertical . ........................ 10dB/div
Referencelevel................... 0dB/div
RFattenuation . .................. 20dB
Videofilter . ..................... 10kHz
ResolutionBW . .................. 1MHz