Performance Verification
AWG510 & AWG520 Service Manual 4-13
. These procedures extend the confidence level provided by the basic
procedures described on page 4–3. The basic procedures should be done first,
then these procedures performed if desired.
The tests in this section comprise an extensive, valid confirmation of perfor-
mance and functionality when the following requirements are met:
HThe cabinet must be installed on the instrument.
HYou must have performed and passed the procedures under Self Tests, found
on page 4–3.
HThe waveform generator must have been last adjusted at an ambient
temperature between +20_ C and +30_ C, must have been operating for a
warm-up period of at least 20 minutes, and must be operating at an ambient
temperature between +10_ C and +40_ C.
Read Conventions on page 4–1.
These procedures use external, traceable signal sources to directly check
warranted characteristics. The required equipment list follows this introduction.
Table4-5: Test equipments
Item number and
description Minimumrequirements Example(recommended) Purpose
1. FrequencyCounter 1 MHz to 100 MHz,
Accuracy: t0.2 ppm
Anritu MF1603A Used to check reference input
test.
2. Digitalmulti meter DC volts range: 0.05 V to 10 V,
Accuracy: "0.1 %
Fluke 8842A Used to check to measure
voltage.
3. Oscilloscope Bandwidth:u500 MHz, 1 MWand
50 Winputs
TektronixTDS784C Checks output signals. Used
in many procedures.
4. Oscilloscope Bandwidth:u1 GHz,
Rise time: t350 ps, 50 Winput
TektronixTDS820 Checks direct DA rise time.
5. SpectrumAnalizer 1 KHz to 1 GHz Tektronix497P or
AdvantestR4131
Checks output signals.
6. FunctionGenerator Output voltage: -5 V to +5 V,
Frequency accuracy: t0.01 %
TektronixAFG310 Generates external input
signals. Used in many input
signal test procedur es.
7. SignalGenerator Output voltage: u0.25 Vp-p,
Frequency: u1 GHz
HP 8648x , x: B or D Used to check external clock
input.
8. BNCCoaxial Cable
(3 required)
50 W,male to male BNC connectors Tektronixpart number
012Ć0482Ć00
Signalinterconnection
PrerequisitesRelated InformationEquipment Required