Performance Tests

AWG510 & AWG520 Service Manual 4-19

AWG510 & AWG520Test Record (Page 1 of 4)
Serial Number: Certificate Number:
Calibration Date: Technician:
Clock frequency tests Lowlimit Test result Highlimit
10 MHz clock frequency accuracy 9,999,990 Hz 10,000,010 Hz
Amplitude and offset tests Low limit Testresult High limit
CH1
Amplit deacc rac
20 mV 17.7 mV 22.3 mV
A
mp
lit
u
d
eaccuracy 200 mV 195 mV 205 mV
2V 1.968 V 2.032 V
CH2 or CH1
Amplit deacc rac
20 mV 17.7 mV 22.3 mV
A
mp
lit
u
d
eaccuracy 200 mV 195 mV 205 mV
2V 1.968 V 2.032 V
CH1
Offsetacc rac
0V -10 mV +10mV
Off
se
t
accuracy 1V 0.980 V 1.020V
-1 V -1.020 V -0.980 V
CH2 or CH1
Offsetacc rac
0V -10 mV +10mV
Off
se
t
accuracy 1V 0.980 V 1.020V
-1 V -1.020 V -0.980 V
Direct DA Output tests Low limit Testresult Highlimit
CH1 Amplitude accuracy 0.45 V 0.55 V
CH2 or CH1 Amplitude accuracy 0.45 V 0.55 V
CH1 Offset accuracy -0.297 V -0.243V
CH2 or CH1 Offset accuracy -0.297 V -0.243V
CH1 Pulse rise time 700 ps
CH2 or CH1 Pulse rise time 700 ps