Performance Verification
 AWG510 & AWG520 Service Manual
d. Set the function generator (AFG310) controls:
   
  

  
   
   
  
2. Set AWG500–Series Waveform Generator controls:
a. Initialize the AWG500–Series Waveform Generator controls:
Push UTILITY (front–panel)System (bottom)
Factory Reset (side)OK (side).
b. Set the amplitude settings:
Push VERTICAL MENU (front–panel)Amplitude (side).
Push 0, ., 0, 2 and ENTER keys in order (0.02).
c. Connect the ADD IN connector to CH1 output:
Push ADD/Direct Out... (side) External (side).
d. Select the waveform file: Load the DC0.WFM as referring to the
procedures on page 4–15.
3. Turn on the AWG500–Series Waveform Generator CH1 output: Push the
RUN and CH1 OUT buttons so that the LEDs above the RUN button and
CH1 output connector light.
4. Check the output level corresponds to the input level:
a. Check the level when the function generator output is off:
Check the reading on DMM: Check that the value is within the range
from –10 mV to +10 mV, and note this reading value.
b. Enable function generator output: Turn on the function generator output.
c. Check the level when the function generator output is on:
Check the reading on DMM: Check that the reading of the DMM
minus the value read in step 4.a is within the range between 0.95 V
and 1.05 V.
5. End procedure: Turn off the function generator output, and disconnect the
function generator and DMM.