Performance Verification
4-4 AWG510 & AWG520 Service Manual
The list on the screen shows the test items and results in the calibration
and diagnostics previously made. For the diagnostics, in addition to
selecting all of the test items shown on the screen, you can select only a
test item that you want to run using the general purpose knob. The result
of the diagnostics are shown as error code. Pass means that the tests have
been made without error. If an error is detected, error code is displayed.
You can also specify how many times the diagnostic tests are performed.
Push the Cycle side button and then turn the general purpose knob to
select the cycle from 1, 3, 10, 100 or Infinite. When you select Infinite,
the tests are repeatedly performed, and are not be terminated until you
push the Abort Diagnostic side button.
Figure 4-1: Diagnostic menu
b. Run the diagnostics: Push the Execute Diagnostic side button. This
executes all the AWG500 diagnostics automatically.
c. Wait: The internal diagnostics do an extensive verification of AWG500
functions. While this verification progresses, the screen displays the
clock icon. When finished, the resulting status appears on the screen.
d. Confirm that no failures are found: Verify that no failures are found and
reported on-screen. If the diagnostics terminates without error, Pass is
displayed instead of the - - -. If a value is displayed, meaning an error is
detected, consult a qualified service technician for further assistance.