Performance Verification
AWG510 & AWG520 Service Manual
2. Set the AWG500–Series Waveform Generator controls and select the
waveform file:
a. Initialize the AWG500–Series Waveform Generator controls:
Push UTILITY (front–panel)
System (bottom)Factory Reset (side)OK (side).
b. Select the waveform for the CH1: Load the MODE.WFM as referring to
the procedures on page 4–15.
3. Turn on the AWG500–Series Waveform Generator output: Push the RUN
button so that the RUN LED lights.
4. Check the CH1 Marker 1 delay function works:
a. Push HORIZONTAL MENU (front–panel)Marker 1 Delay (side)
b. Continuously change the Marker 1 delay from 0 s to 2.0 ns by turning
the general purpose knob.
c. Confirm that the Marker 1 output delay relative to the Marker 2 output
varies from 0 s to 2.0 ns on the oscilloscope screen.
5. Set the AWG500–Series Waveform Generator and oscilloscope controls:
a. Enter numeric value of 0 s: Push 0 and ENTER keys to set the Marker 1
Delay back to 0 s.
b. Change the oscilloscope controls: Change the oscilloscope trigger source
from CH2 to CH1.
6. Check the CH1 Marker 2 delay function works:
a. Push the Marker 2 Delay side button.
b. Continuously change the Marker 2 delay from 0 s to 2.0 ns by turning
the general purpose knob.
c. Confirm that the Marker 2 output delay relative to the Marker 1 output
varies from 0 s to about 2.0 ns on the oscilloscope screen.