Performance Verification
AWG510 & AWG520 Service Manual 4-57
1. Use the test hookup and oscilloscope settings from previous check.
2. Set the AWG500–Series Waveform Generator controls and select the
sequence file:
a. Initialize the AWG500–Series Waveform Generator controls:
Push UTILITY (front–panel)
System (bottom)Factory Reset (side)OK (side).
b. Select the waveform file: Load the PT_STROB.SEQ as referring to the
procedures on page 4–15.
c. Set the AWG500–Series Waveform Generator to enhanced mode: Push
SETUP (front–panel)Run Mode (bottom)Enhanced (side) to set
the run mode to enhanced.
3. Turn on the AWG500–Series Waveform Generator CH1 output: Push the
RUN and CH1 OUT buttons so that the LEDs above the RUN button and
CH1 output connector light.
4. Check the EVENT IN connector strobe pin input:
a. Confirm the waveform on the oscilloscope: Confirm that the waveform
being displayed on the oscilloscope is as shown in Figure 4–27.
Figure 4-27: Initial waveformoutput
b. Generate a strobe signal: Close the SW5 of the ground closure to
generate an event signal on the EVENT IN connector strobe pin.
c. Confirm the waveform on the oscilloscope: Confirm that the oscilloscope
displays the dc waveform as shown in Figure 4–28.
Check Strobe Input