Performance Verification
4-58 AWG510 & AWG520 Service Manual
Figure 4-28: DC waveform output when theSW5 is closed
d. Degenerate the strobe signal: Open the SW5 of the ground closure to
degenerate the strobe signal on the EVENT IN connector strobe pin.
e. Confirm the waveform on the oscilloscope: Confirm that the dc
waveform is kept displaying on the oscilloscope.
f. Generate a strobe signal: Close the SW5 of the ground closure again.
g. Confirm the waveform on the oscilloscope: Confirm that the oscilloscope
displays back the waveform in Figure 4–27.
5. End procedure: Disconnect the oscilloscope and ground closure.