Performance Verification
4-24 AWG510 & AWG520 Service Manual
2. Set the AWG500–Series Waveform Generator controls and select the
waveform file:
a. Initialize the AWG500–Series Waveform Generator controls: Push
UTILITY (front–panel)
System (bottom)Factory Reset (side)OK (side).
b. Select the waveform file: Load the MODE.WFM as referring to the
procedures on page 4–15.
3. Turn on the AWG500–Series Waveform Generator CH1 output: Push the
RUN and CH1 OUT buttons so that the LEDs above the RUN button and
CH1 output connector light.
4. Check against limits: Check that the amplitude of the sine wave displayed on
the oscilloscope is 5 vertical divisions and that the waveform of one cycle
per 2 horizontal divisions is displayed.
5. End procedure: Disconnect the oscilloscope.
Equipment
required
! '& "   #! $# !#!
  ""
Prerequisites  !"%! !#!#" #!! $"#"
"#  
1. Install the test hookup and set test equipment controls:
a. Hook up the oscilloscope: Connect the AWG500–Series Waveform
Generator CH1 output connector through the 50 W BNC coaxial cable to
the oscilloscope CH1 input connector.
b. Hook up the function generator: Connect the AWG500–Series Wave-
form Generator TRIG IN connector though a 50 W BNC coaxial cable
and BNC–T connector to the function generator output connector, and
connect a 50 W BNC coaxial cable to BNC–T connector and the
oscilloscope CH2 input (see Figure 4–6).
Check Triggered Mode