B&K 8510 instruction manual Enter

Page 33

4) Press Enter and begin to edit the automatic testing file, , VFD shows MAX CURR= .**A, which means setting the maximum of voltage. The maximum is bigger than 3A, which means CC mode is in high value, here, the maximum is 3A. Press 3 and press Enter to confirm.

5)VFD displays MAX VOLT=.**V, which means setting the maximum of voltage, the maximum is bigger than

18V, which means CV is in high value, here ,the maximum is 18V. Press 1 + 8 to set MAX VOLT=18V,

press

Enter

to confirm.

 

6)VFD displays MAX POWER=.**W, which means setting the maximum of power, here, it is150W,

Press 1 + 5 + 0 to set MAX POWER=150W, press Enter to confirm.

7)VFD displays TEST COUNT=**, which means setting steps of testing,20 steps are permitted at most, here,

they are 6 steps. Press 6 to set TEST COUNT=6, press Enter to confirm.

8)Setting the mode of current procedure, press ▲▼to choose one mode in CONST CURRENTCONST VOLTAGECONST POWERCONST RESIS. Here, the first step is CC mode, press ▲▼to >CONST CURRENT, press Enter to confirm.

9)VFD shows SET**=.**, which means setting the current procedure. Here, the first sis 0.21A, press 0

 

2 1 , then press Enter to confirm.

 

10)

Setting whether the current procedure is short circuit or not, press ▲▼to choose one in SHORT ON or

 

SHORT OFF. Here the first step is SHORT OFF. Press SHORT OFF, press Enter

to confirm.

11)

Select the test number of this step, press▲▼ to choose >READBACK V, PRESS

Enter to confirm.

12)VFD displays MIN**=.**, which means setting the lower limit. Here, the first step is 5.8V. Press 5

8 , press Enter to confirm.

13)VFD displays MAX**=.**, which requests setting the upper limit. Here, the first step is 6.15V, press

61 5 ,then press Enter to confirm.

14)VFD shows DELAY**=.**, which means setting the prolong time, that is how long we can read the testing value after inputting the value we settled. To wait for testing after the input stable, the range of prolong time is 025.5 seconds, when the setting is 25.5 seconds, automatic testing will be pause, only the user

press

Shift

+ Trigger ,the testing will continue. Here, the first step is 1 second, press 1 and then

press

Enter

to confirm.

33

Image 33
Contents Instruction Manual Programmable DC Electronic Load Table of Contents Warranty Information Safety Symbols Front Panel Quick ReferenceFront Panel Display Rear PanelFront Panel Menus Immediate Action KeysKey Pad TranMenu Operation POWER-ON RecallADC Update Rate High Lowdefault General Information Options and AccessoriesDocument Orientation Getting Started MapAccessories Features and CapabilitiesDescription Front Panel ControlsConstant Voltage CV Mode Constant Resistance CR ModeRemote Programming Operating ModesConstant Power CW Mode Transient OperationContinuous Trigger Mode List OperationPulse Triggered Operation Input ControlShort On/Off Input On/Off Protection FeaturesOperation Range Over VoltageOver Temperature Over PowerPin trigger and remote sensing connectors Remote Sense Function and Trigger FunctionWiring diagram for remote sense Saving and Recalling SettingsOperation Battery TestingAction for set Von/Voff value Von/Voff OperationInstallation InspectionLocation Outline Diagram Power CordBench Operation Input ConnectionsTurn-On Checkout Checkout ProcedureIntroduction Trigger and Remote Sensing ConnectionsCase of Trouble Front Panel OperationDisplay Front Panel DisplaySet set up a constant current from 0 to Max current PowerExample Function keysSet set up a constant resistance from 0.1Ω to 4000Ω Set set up a constant power from 0 to Max powerPOWER=*.***W 000VSet set up a constant voltage from 1.5V to Max voltage Transition Testing OperationOn/Off input setting Shift + StoreTransition Parameter Setting Continuous Transient OperationPulse Transient Operation Steps Toggled Transient OperationList Operation Automatic Testing Function of Electronic Load Edit the Testing FileQuick Recall Function Enter Setting the Resolution and number of decimal points Automatic TestingRange Accuracy Resolution SpecificationsCommunication cable Remote Operation ModeIT-E131 RS232 Communication cable IT-E132 USB Communication cableCommunication between electronic load and PC Frame FormatIT-E132 communication 2AH Communication Protocol Selecting the Remote control mode(20H)Setting / Reading max input voltage(22H/23H) Setting / Reading the max input current .(24H/25H)Selecting the input on/off state (21H) Setting / Reading max input power(26H/27H) Setting / Reading CC mode current value(2AH/2BH) Setting / Reading CV mode voltage value.(2CH/2DH)Setting / Reading CW mode watt value(2EH/2FH) Setting / Reading CR mode resistance value(30H/31H) 26th byte Sum code Setting / Reading the number of list steps EH/3FH Setting /Reading the list repeat mode CH/3DHSetting / Reading List file name 48H/49H Save / Get list file in appointed area. CH/4DH Setting / Reading timer value of for Load on 50H/51H)Setting communication address 54H Enable/Disable Local control mode HSelecting / Reading trigger source H/59H Reading input voltage, current, power and relative state FH Sending a trigger signal to trigging the electronic load AHBIT Enter the calibration mode(60H)Calibrate voltage value(62H) Getting the calibration mode state(61H)Setting / Reading calibration information(67H/68H) Calibrate current value(64H)Store the calibration data to EEPROM(66H) Restore the factory default calibration data(69H) For example:Reading information in bar code 6BH Setting information of bar code(6CH) Limited One-Year Warranty Service InformationDeclaration of CE Conformity Savi Ranch Parkway Yorba Linda, CA