B&K 8510 instruction manual Setting / Reading max input power(26H/27H)

Page 41

3rd byte

Command24H/25H

 

 

 

4th byte

The Lowest byte of max current value

5th byte

The Lowest byte of max current value

6th byte

The higher byte of max current value

7th byte

The highest byte of max current value

From 8th to 25th byte

System reserve

26th bye

Sum code

NOTE

Represent an current value by 4 bytes of Hex .Lower bytes are in the front location, higher bytes are in the later location.1 represent 0.1mA,If setting upper limit is 3.0000Athe hex code is 0X00007530then the 4th byte is 0X305th is 0X756th is 0X007th is 0X00

5. Setting / Reading max input power26H/27H

1st byte

Start bit ( AAH )

2nd byte

Address (0—0XFE)

3rd byte

Command26H/27H

 

 

 

4th byte

The lowest byte of max power value.

5th byte

The lower byte of max power value

6th byte

The higher byte of max power value.

7th byte

The highest byte of max power value.

From 8th to 25th byte

System reserve

26th byte

Sum code

NOTE

Represent power value by 4 bytes of Hex. Lower bytes are in the

Front location, higher bytes are in the later location. 1 represents 1mW. If setting upper value is 200.000Wthe hex code is 0X00030d40then the 4th byte is

0X405th is 0X0d6th is 0X037th is 0X00.

6. Selecting / Reading CC/CV/CW/CR operation mode of Electronic load. 28H/29H

1st byte

Start bit ( AAH )

2nd byte

Address (0—0XFE)

3rd byte

Command28H/29H

 

 

 

4th byte

Mode0 is CC mode, 1 is CV mode , 2 is CW mode , 3 is CR mode

 

 

 

From 5th to 25th byte

System reserve

26th byte

Sum code

41

Image 41
Contents Instruction Manual Programmable DC Electronic Load Table of Contents Warranty Information Safety Symbols Front Panel Quick ReferenceFront Panel Display Rear PanelFront Panel Menus Immediate Action KeysKey Pad TranMenu Operation POWER-ON RecallADC Update Rate High Lowdefault General Information Options and AccessoriesDocument Orientation Getting Started MapAccessories Features and CapabilitiesDescription Front Panel ControlsConstant Voltage CV Mode Constant Resistance CR ModeRemote Programming Operating ModesContinuous Constant Power CW ModeTransient Operation Pulse Trigger ModeList Operation Short On/Off Triggered OperationInput Control Input On/Off Protection FeaturesOperation Range Over VoltageOver Temperature Over PowerPin trigger and remote sensing connectors Remote Sense Function and Trigger FunctionWiring diagram for remote sense Saving and Recalling SettingsOperation Battery TestingAction for set Von/Voff value Von/Voff OperationLocation InstallationInspection Outline Diagram Power CordBench Operation Input ConnectionsTurn-On Checkout Checkout ProcedureIntroduction Trigger and Remote Sensing ConnectionsCase of Trouble Front Panel OperationDisplay Front Panel DisplaySet set up a constant current from 0 to Max current PowerExample Function keysSet set up a constant resistance from 0.1Ω to 4000Ω Set set up a constant power from 0 to Max powerPOWER=*.***W 000VSet set up a constant voltage from 1.5V to Max voltage Transition Testing OperationOn/Off input setting Shift + StorePulse Transient Operation Transition Parameter SettingContinuous Transient Operation Steps Toggled Transient OperationList Operation Quick Recall Function Automatic Testing Function of Electronic LoadEdit the Testing File Enter Setting the Resolution and number of decimal points Automatic TestingRange Accuracy Resolution SpecificationsCommunication cable Remote Operation ModeIT-E131 RS232 Communication cable IT-E132 USB Communication cableIT-E132 communication Communication between electronic load and PCFrame Format 2AH Communication Protocol Selecting the Remote control mode(20H)Selecting the input on/off state (21H) Setting / Reading max input voltage(22H/23H)Setting / Reading the max input current .(24H/25H) Setting / Reading max input power(26H/27H) Setting / Reading CW mode watt value(2EH/2FH) Setting / Reading CC mode current value(2AH/2BH)Setting / Reading CV mode voltage value.(2CH/2DH) Setting / Reading CR mode resistance value(30H/31H) 26th byte Sum code Setting / Reading the number of list steps EH/3FH Setting /Reading the list repeat mode CH/3DHSetting / Reading List file name 48H/49H Save / Get list file in appointed area. CH/4DH Setting / Reading timer value of for Load on 50H/51H)Selecting / Reading trigger source H/59H Setting communication address 54HEnable/Disable Local control mode H Reading input voltage, current, power and relative state FH Sending a trigger signal to trigging the electronic load AHBIT Enter the calibration mode(60H)Calibrate voltage value(62H) Getting the calibration mode state(61H)Store the calibration data to EEPROM(66H) Setting / Reading calibration information(67H/68H)Calibrate current value(64H) Reading information in bar code 6BH Restore the factory default calibration data(69H)For example: Setting information of bar code(6CH) Limited One-Year Warranty Service InformationDeclaration of CE Conformity Savi Ranch Parkway Yorba Linda, CA