B&K 8510 instruction manual Automatic Testing, Setting the Resolution and number of decimal points

Page 34

15)Repeat 9)14), and set the other process step by step as follows:

A.CONST CURRENT,0A,SHORT OFF, READ BACK V, 5.9V, 6.4V,1S

B.CONST VOLTAGE, 5V, SHORT OFF,READ BACK A, 0.205A, 0.245A,1S

C.CONST VOLTAGE, 3V, SHORT OFF, READ BACK A, 0.205A,0.245A,1S

D.CONST VOLTAGE, 2V,SHORT OFF, READ BACK A, 0.205A,0.245A, 1S

E.CONST CURRENT, 0A, SHORT ON, READ BACK A,0A,0.245A, 1S

16)VFD displays SHORT TEST FILE*, which requests saving the files edited to EEPROM, automatically testing files and LIST files use a common memory area, and the area can contain 8 groups of automatic testing files for

next usage. Here, the testing file could be saved in the first group, press 1 and press Enter to confirm.

17) When the edition of automatic testing file is finished, press Enter two times to back the menu.

Taking out the Testing File Rapidly

1)

Press

Shift

+ Menu to enter into menu, VFD displays>CONFIG

2) Press▼ and move the menu to>LIST SET

3) Press

Enter

to enter into submenu, VFD displays>MODE SET

4)

Press▼ and move the menu to>CALL TEST FILE

5)

Press

Enter

and recall the file edited

Automatic Testing

After editing the automatic testing file, the automatic testing could be carried out by the below steps.

1) Press

Shift

+ I-set

to enter into automatic testing, VFD displays the name of current file: NAME: TEST

FILE1

 

 

 

2) Press

Enter

to observe the voltage and current we inputted, and also can observe the step number of testing.

3) Press

Shift

+ Trigger

and begin to operate, when the automatic testing is pause (the current prolong time of

one step is 25.5S), press

Shift + Trigger also can make the testing continue.

4)When the automatic testing is finished, the buzzer will hint by tweet (buzzer), VFD reads the result of testing. If the testing passed, VFD reads PASS, or reads FAULT. Press▲▼ also to observe the numerical value and result of every step.

5)Press ESC to back the automatic testing.

Setting the Resolution and number of decimal points

By selecting the low range and high range, you can measure the voltage and current more precisely.

34

Image 34
Contents Instruction Manual Programmable DC Electronic Load Table of Contents Warranty Information Safety Symbols Quick Reference Front PanelRear Panel Front Panel DisplayKey Pad Immediate Action KeysFront Panel Menus TranPOWER-ON Recall Menu OperationADC Update Rate High Lowdefault Document Orientation Options and AccessoriesGeneral Information Getting Started MapDescription Features and CapabilitiesAccessories Front Panel ControlsRemote Programming Constant Resistance CR ModeConstant Voltage CV Mode Operating ModesTransient Operation Constant Power CW ModeContinuous List Operation Trigger ModePulse Input Control Triggered OperationShort On/Off Operation Range Protection FeaturesInput On/Off Over VoltagePin trigger and remote sensing connectors Over PowerOver Temperature Remote Sense Function and Trigger FunctionSaving and Recalling Settings Wiring diagram for remote senseBattery Testing OperationVon/Voff Operation Action for set Von/Voff valueInspection InstallationLocation Bench Operation Power CordOutline Diagram Input ConnectionsIntroduction Checkout ProcedureTurn-On Checkout Trigger and Remote Sensing ConnectionsDisplay Front Panel OperationCase of Trouble Front Panel DisplayExample PowerSet set up a constant current from 0 to Max current Function keysPOWER=*.***W Set set up a constant power from 0 to Max powerSet set up a constant resistance from 0.1Ω to 4000Ω 000VOn/Off input setting Transition Testing OperationSet set up a constant voltage from 1.5V to Max voltage Shift + StoreContinuous Transient Operation Transition Parameter SettingPulse Transient Operation Toggled Transient Operation StepsList Operation Edit the Testing File Automatic Testing Function of Electronic LoadQuick Recall Function Enter Automatic Testing Setting the Resolution and number of decimal pointsSpecifications Range Accuracy ResolutionIT-E131 RS232 Communication cable Remote Operation ModeCommunication cable IT-E132 USB Communication cableFrame Format Communication between electronic load and PCIT-E132 communication 2AH Selecting the Remote control mode(20H) Communication ProtocolSetting / Reading the max input current .(24H/25H) Setting / Reading max input voltage(22H/23H)Selecting the input on/off state (21H) Setting / Reading max input power(26H/27H) Setting / Reading CV mode voltage value.(2CH/2DH) Setting / Reading CC mode current value(2AH/2BH)Setting / Reading CW mode watt value(2EH/2FH) Setting / Reading CR mode resistance value(30H/31H) 26th byte Sum code Setting /Reading the list repeat mode CH/3DH Setting / Reading the number of list steps EH/3FHSetting / Reading List file name 48H/49H Setting / Reading timer value of for Load on 50H/51H) Save / Get list file in appointed area. CH/4DHEnable/Disable Local control mode H Setting communication address 54HSelecting / Reading trigger source H/59H Sending a trigger signal to trigging the electronic load AH Reading input voltage, current, power and relative state FHEnter the calibration mode(60H) BITGetting the calibration mode state(61H) Calibrate voltage value(62H)Calibrate current value(64H) Setting / Reading calibration information(67H/68H)Store the calibration data to EEPROM(66H) For example: Restore the factory default calibration data(69H)Reading information in bar code 6BH Setting information of bar code(6CH) Service Information Limited One-Year WarrantyDeclaration of CE Conformity Savi Ranch Parkway Yorba Linda, CA