B&K 8510 instruction manual Getting the calibration mode state(61H), Calibrate voltage value(62H)

Page 51

If Load is not in protection state, users could do the calibration operation.

37. Getting the calibration mode state61H

1st byte

Start bit ( AAH )

2nd byte

Address (0—0XFE)

3rd byte

Command61H

 

 

 

 

 

 

4th byte

Calibration protection state

From 5th to 25th byte

System

26th byte

Sum code

 

 

 

 

NOTE

Represent calibration protection state by one byte. Each byte is defined as:

From high to low

7 6 5 4

3 2 1 0

0 byteprotection state0 represent not in the protection state 1 represent in protection state.

38. Calibrate voltage value62H

1st byte

Start bit ( AAH )

2nd byte

Address (0—0XFE)

3rd byte

Command62H

 

 

 

 

 

 

4th byte

Voltage calibration point1~4

 

 

 

 

 

 

From 5th to 25th byte

System reserve

26th byte

Sum code

 

 

 

 

NOTE

Current calibration standard points have four: 1, 2, 3,4.

39. Sending the actual input voltage to calibration program63H

1st byte

Start bit ( AAH )

2nd byte

Address (0—0XFE)

3rd byte

Command63H

 

 

 

4th byte

The lowest byte of actual voltage

5th byte

The lower byte of actual voltage

6th byte

The higher byte of actual voltage.

7th byte

The highest byte of actual voltage.

From 8th to 25th byte

System reserve

26th byte

Sum code

51

Image 51
Contents Instruction Manual Programmable DC Electronic Load Table of Contents Warranty Information Safety Symbols Front Panel Quick ReferenceFront Panel Display Rear PanelTran Immediate Action KeysFront Panel Menus Key PadMenu Operation POWER-ON RecallADC Update Rate High Lowdefault Getting Started Map Options and AccessoriesGeneral Information Document OrientationFront Panel Controls Features and CapabilitiesAccessories DescriptionOperating Modes Constant Resistance CR ModeConstant Voltage CV Mode Remote ProgrammingConstant Power CW Mode Transient OperationContinuous Trigger Mode List OperationPulse Triggered Operation Input ControlShort On/Off Over Voltage Protection FeaturesInput On/Off Operation RangeRemote Sense Function and Trigger Function Over PowerOver Temperature Pin trigger and remote sensing connectorsWiring diagram for remote sense Saving and Recalling SettingsOperation Battery TestingAction for set Von/Voff value Von/Voff OperationInstallation InspectionLocation Input Connections Power CordOutline Diagram Bench OperationTrigger and Remote Sensing Connections Checkout ProcedureTurn-On Checkout IntroductionFront Panel Display Front Panel OperationCase of Trouble DisplayFunction keys PowerSet set up a constant current from 0 to Max current Example000V Set set up a constant power from 0 to Max powerSet set up a constant resistance from 0.1Ω to 4000Ω POWER=*.***WShift + Store Transition Testing OperationSet set up a constant voltage from 1.5V to Max voltage On/Off input settingTransition Parameter Setting Continuous Transient OperationPulse Transient Operation Steps Toggled Transient OperationList Operation Automatic Testing Function of Electronic Load Edit the Testing FileQuick Recall Function Enter Setting the Resolution and number of decimal points Automatic TestingRange Accuracy Resolution SpecificationsIT-E132 USB Communication cable Remote Operation ModeCommunication cable IT-E131 RS232 Communication cableCommunication between electronic load and PC Frame FormatIT-E132 communication 2AH Communication Protocol Selecting the Remote control mode(20H)Setting / Reading max input voltage(22H/23H) Setting / Reading the max input current .(24H/25H)Selecting the input on/off state (21H) Setting / Reading max input power(26H/27H) Setting / Reading CC mode current value(2AH/2BH) Setting / Reading CV mode voltage value.(2CH/2DH)Setting / Reading CW mode watt value(2EH/2FH) Setting / Reading CR mode resistance value(30H/31H) 26th byte Sum code Setting / Reading the number of list steps EH/3FH Setting /Reading the list repeat mode CH/3DHSetting / Reading List file name 48H/49H Save / Get list file in appointed area. CH/4DH Setting / Reading timer value of for Load on 50H/51H)Setting communication address 54H Enable/Disable Local control mode HSelecting / Reading trigger source H/59H Reading input voltage, current, power and relative state FH Sending a trigger signal to trigging the electronic load AHBIT Enter the calibration mode(60H)Calibrate voltage value(62H) Getting the calibration mode state(61H)Setting / Reading calibration information(67H/68H) Calibrate current value(64H)Store the calibration data to EEPROM(66H) Restore the factory default calibration data(69H) For example:Reading information in bar code 6BH Setting information of bar code(6CH) Limited One-Year Warranty Service InformationDeclaration of CE Conformity Savi Ranch Parkway Yorba Linda, CA