B&K 8510 instruction manual Front Panel Operation, Case of Trouble, Front Panel Display

Page 25

6.Press △▽

0.000W I: 0.000A

Display the actual power value and setting value.

In Case of Trouble

Should the power-on self test of the Electronic Load fail, you may try the following: When turning on the power, make sure the power cord to the unit is plugged in and the power switch is in the on position.

1)Check the power voltage setting.

Work voltage of load have two type 110V or 220V , Please make sure it is right voltage accordance to the voltage in your area. You could change the voltage setting through dial the switch in the rear panel.

2)Check the fuse.

If fuse is blown, please replace it according to the following specification.

Model

Fuse specification

Fuse specification

 

(110VAC)

(220VAC)

8510

T0.5A

250V

T0.3A

250V

3) Location of Fuse

Fuse

Front Panel Operation

Display

16-character fluorescent display for showing measurements and programmed values.

Front Panel Display

Indicators for operating modes and status conditions:

OFF

power off

Trigger

Indicates that the electronic load is waiting an

 

 

 

initiate and trigger to occur.

 

 

 

 

CC

The selected input channel is in the

Sense

Indicates that the electronic load is in sense

 

constant current (CC) mode.

 

state

 

 

 

 

CV

The selected input channel is in the

Error

A errors have occurred

 

constant voltage (CV) mode.

 

 

 

 

 

 

CW

The selected input channel is in the

Link

In the communication state

 

constant power (CW) mode.

 

 

 

 

 

 

25

Image 25
Contents Instruction Manual Programmable DC Electronic Load Table of Contents Warranty Information Safety Symbols Front Panel Quick ReferenceFront Panel Display Rear PanelFront Panel Menus Immediate Action KeysKey Pad TranMenu Operation POWER-ON RecallADC Update Rate High Lowdefault General Information Options and AccessoriesDocument Orientation Getting Started MapAccessories Features and CapabilitiesDescription Front Panel ControlsConstant Voltage CV Mode Constant Resistance CR ModeRemote Programming Operating ModesTransient Operation Constant Power CW ModeContinuous List Operation Trigger ModePulse Input Control Triggered OperationShort On/Off Input On/Off Protection FeaturesOperation Range Over VoltageOver Temperature Over PowerPin trigger and remote sensing connectors Remote Sense Function and Trigger FunctionWiring diagram for remote sense Saving and Recalling SettingsOperation Battery TestingAction for set Von/Voff value Von/Voff OperationInspection InstallationLocation Outline Diagram Power CordBench Operation Input ConnectionsTurn-On Checkout Checkout ProcedureIntroduction Trigger and Remote Sensing ConnectionsCase of Trouble Front Panel OperationDisplay Front Panel DisplaySet set up a constant current from 0 to Max current PowerExample Function keysSet set up a constant resistance from 0.1Ω to 4000Ω Set set up a constant power from 0 to Max powerPOWER=*.***W 000VSet set up a constant voltage from 1.5V to Max voltage Transition Testing OperationOn/Off input setting Shift + StoreContinuous Transient Operation Transition Parameter SettingPulse Transient Operation Steps Toggled Transient OperationList Operation Edit the Testing File Automatic Testing Function of Electronic LoadQuick Recall Function Enter Setting the Resolution and number of decimal points Automatic TestingRange Accuracy Resolution SpecificationsCommunication cable Remote Operation ModeIT-E131 RS232 Communication cable IT-E132 USB Communication cableFrame Format Communication between electronic load and PCIT-E132 communication 2AH Communication Protocol Selecting the Remote control mode(20H)Setting / Reading the max input current .(24H/25H) Setting / Reading max input voltage(22H/23H)Selecting the input on/off state (21H) Setting / Reading max input power(26H/27H) Setting / Reading CV mode voltage value.(2CH/2DH) Setting / Reading CC mode current value(2AH/2BH)Setting / Reading CW mode watt value(2EH/2FH) Setting / Reading CR mode resistance value(30H/31H) 26th byte Sum code Setting / Reading the number of list steps EH/3FH Setting /Reading the list repeat mode CH/3DHSetting / Reading List file name 48H/49H Save / Get list file in appointed area. CH/4DH Setting / Reading timer value of for Load on 50H/51H)Enable/Disable Local control mode H Setting communication address 54HSelecting / Reading trigger source H/59H Reading input voltage, current, power and relative state FH Sending a trigger signal to trigging the electronic load AHBIT Enter the calibration mode(60H)Calibrate voltage value(62H) Getting the calibration mode state(61H)Calibrate current value(64H) Setting / Reading calibration information(67H/68H)Store the calibration data to EEPROM(66H) For example: Restore the factory default calibration data(69H)Reading information in bar code 6BH Setting information of bar code(6CH) Limited One-Year Warranty Service InformationDeclaration of CE Conformity Savi Ranch Parkway Yorba Linda, CA