B&K 8510 Calibrate current value(64H), Store the calibration data to EEPROM(66H)

Page 52

40. Calibrate current value64H

1st byte

Start bit ( AAH )

2nd byte

Address (0—0XFE)

3rd byte

Command64H

 

 

 

 

 

 

4th byte

Current calibration point 1~4

 

 

 

 

 

 

From 5th to 25th byte

System reserve

26th byte

Sum code

 

 

 

 

NOTE

Current calibration standard points have four: 1,2,3,4

41. Sending the actual input current to calibration program (65H

1st byte

Start bit ( AAH )

2nd byte

Address (0—0XFE)

3rd byte

Command65H

 

 

 

4th byte

The lowest byte of actual current

5th byte

The lower byte of actual current

6th byte

The higher byte of actual current

7th byte

The highest byte of actual current

From 8th to 25th byte

System reserve

26th byte

Sum code

42. Store the calibration data to EEPROM66H

1st byte

Start bit ( AAH )

2nd byte

Address (0—0XFE)

3rd byte

Command 66H

 

 

 

 

 

 

From 4th to 25th byte

System reserve

26th byte

Sum code

 

 

 

 

NOTE

Finish the calibration operation, users should save the calibration parameter in EEPROM with this command; users could use these data in next power on.

43. Setting / Reading calibration information67H/68H

1st byte

2nd byte

Start bit ( AAH )

Address (0—0XFE)

52

Image 52
Contents Instruction Manual Programmable DC Electronic Load Table of Contents Warranty Information Safety Symbols Quick Reference Front PanelRear Panel Front Panel DisplayImmediate Action Keys Front Panel MenusKey Pad TranPOWER-ON Recall Menu OperationADC Update Rate High Lowdefault Options and Accessories General InformationDocument Orientation Getting Started MapFeatures and Capabilities AccessoriesDescription Front Panel ControlsConstant Resistance CR Mode Constant Voltage CV ModeRemote Programming Operating ModesTransient Operation Constant Power CW ModeContinuous List Operation Trigger ModePulse Input Control Triggered OperationShort On/Off Protection Features Input On/OffOperation Range Over VoltageOver Power Over TemperaturePin trigger and remote sensing connectors Remote Sense Function and Trigger FunctionSaving and Recalling Settings Wiring diagram for remote senseBattery Testing OperationVon/Voff Operation Action for set Von/Voff valueInspection InstallationLocation Power Cord Outline DiagramBench Operation Input ConnectionsCheckout Procedure Turn-On CheckoutIntroduction Trigger and Remote Sensing ConnectionsFront Panel Operation Case of TroubleDisplay Front Panel DisplayPower Set set up a constant current from 0 to Max currentExample Function keysSet set up a constant power from 0 to Max power Set set up a constant resistance from 0.1Ω to 4000ΩPOWER=*.***W 000VTransition Testing Operation Set set up a constant voltage from 1.5V to Max voltageOn/Off input setting Shift + StoreContinuous Transient Operation Transition Parameter SettingPulse Transient Operation Toggled Transient Operation StepsList Operation Edit the Testing File Automatic Testing Function of Electronic LoadQuick Recall Function Enter Automatic Testing Setting the Resolution and number of decimal pointsSpecifications Range Accuracy ResolutionRemote Operation Mode Communication cableIT-E131 RS232 Communication cable IT-E132 USB Communication cableFrame Format Communication between electronic load and PCIT-E132 communication 2AH Selecting the Remote control mode(20H) Communication ProtocolSetting / Reading the max input current .(24H/25H) Setting / Reading max input voltage(22H/23H)Selecting the input on/off state (21H) Setting / Reading max input power(26H/27H) Setting / Reading CV mode voltage value.(2CH/2DH) Setting / Reading CC mode current value(2AH/2BH)Setting / Reading CW mode watt value(2EH/2FH) Setting / Reading CR mode resistance value(30H/31H) 26th byte Sum code Setting /Reading the list repeat mode CH/3DH Setting / Reading the number of list steps EH/3FHSetting / Reading List file name 48H/49H Setting / Reading timer value of for Load on 50H/51H) Save / Get list file in appointed area. CH/4DHEnable/Disable Local control mode H Setting communication address 54HSelecting / Reading trigger source H/59H Sending a trigger signal to trigging the electronic load AH Reading input voltage, current, power and relative state FHEnter the calibration mode(60H) BITGetting the calibration mode state(61H) Calibrate voltage value(62H)Calibrate current value(64H) Setting / Reading calibration information(67H/68H)Store the calibration data to EEPROM(66H) For example: Restore the factory default calibration data(69H)Reading information in bar code 6BH Setting information of bar code(6CH) Service Information Limited One-Year WarrantyDeclaration of CE Conformity Savi Ranch Parkway Yorba Linda, CA