B&K 8510 instruction manual ADC Update Rate High Lowdefault

Page 10

ADC UPDATE RATE

HIGH

LOW<DEFAULT>

TRIGGER SOURCE Choosing the trigger signals source.

 

 

IMMEDIATE<DEF>

Trigger signals from Shift + Trigger key

 

 

EXTERNAL

 

Trigger signals from the TRIG connector in the rear panel.

 

 

BUS

 

Communication command trigger mode.

 

CONNECT MODE

 

 

 

 

 

MAXTIDLEXING

 

 

 

 

SEPARATE<DEF>

 

 

 

BAUDRATE SET

Setting baud rate.

 

 

 

4800<DEFAULT>

 

 

 

 

9600

 

 

 

 

 

19200

 

 

 

 

 

38400

 

 

 

 

COMM. PARITY SET

Command parity setting.

 

 

NONE<DEFAULT>

 

 

 

 

EVEN

 

 

 

 

 

ODD

 

 

 

 

ADDRESS SET

Setting communication Flow mode

 

KEY LOCK SET

Setting keypad password.

 

 

 

Press

Enter

directly to disable the key lock function.

 

EXIT

 

 

 

SYSTTEM SET

 

MAX CURRENT SET

Setup the Maximum current.

 

MAX POWER SET

Setup the Maximum Power.

 

MAX VOLTAGE SET

Setup the Maximum Voltage.

 

VOLTAGE ONSET

 

 

 

VOLTAGE OFFSET

 

 

 

EXIT

 

 

 

LIST SET

 

 

 

 

 

MODE SET

 

Setting operation mode.

 

 

FIXED MODE

 

Fixed mode.

 

 

LIST MODE

 

Choosing List mode.

 

CALL LIST FILE

Recall list operation file.

 

EDIT LIST FILE

 

Edit list operation file.

 

LIST STORE MODE

Users can choose 4 kind of memory space to save the list file.

 

 

8 X 120

STEPS

Total 8 files and each file have120 list steps.

 

 

4 X 250

STEPS

Total 4 files and each file have250 list steps.

 

 

2 X 500

STEPS

Total 2 files and each file have500 list steps.

 

 

1 X 1000 STEPS

Total 1 file and each file have1000 list steps.

 

EXIT

 

 

 

LOAD ON TIMER

 

 

 

 

 

TIMER STATE

 

Setting LOAD ON timer state

10

Image 10
Contents Instruction Manual Programmable DC Electronic Load Table of Contents Warranty Information Safety Symbols Quick Reference Front PanelRear Panel Front Panel DisplayKey Pad Immediate Action KeysFront Panel Menus TranPOWER-ON Recall Menu OperationADC Update Rate High Lowdefault Document Orientation Options and AccessoriesGeneral Information Getting Started MapDescription Features and CapabilitiesAccessories Front Panel ControlsRemote Programming Constant Resistance CR ModeConstant Voltage CV Mode Operating ModesTransient Operation Constant Power CW ModeContinuous List Operation Trigger ModePulse Input Control Triggered OperationShort On/Off Operation Range Protection FeaturesInput On/Off Over VoltagePin trigger and remote sensing connectors Over PowerOver Temperature Remote Sense Function and Trigger FunctionSaving and Recalling Settings Wiring diagram for remote senseBattery Testing OperationVon/Voff Operation Action for set Von/Voff valueInspection InstallationLocation Bench Operation Power CordOutline Diagram Input ConnectionsIntroduction Checkout ProcedureTurn-On Checkout Trigger and Remote Sensing ConnectionsDisplay Front Panel OperationCase of Trouble Front Panel DisplayExample PowerSet set up a constant current from 0 to Max current Function keysPOWER=*.***W Set set up a constant power from 0 to Max powerSet set up a constant resistance from 0.1Ω to 4000Ω 000VOn/Off input setting Transition Testing OperationSet set up a constant voltage from 1.5V to Max voltage Shift + StoreContinuous Transient Operation Transition Parameter SettingPulse Transient Operation Toggled Transient Operation StepsList Operation Edit the Testing File Automatic Testing Function of Electronic LoadQuick Recall Function Enter Automatic Testing Setting the Resolution and number of decimal pointsSpecifications Range Accuracy ResolutionIT-E131 RS232 Communication cable Remote Operation ModeCommunication cable IT-E132 USB Communication cableFrame Format Communication between electronic load and PCIT-E132 communication 2AH Selecting the Remote control mode(20H) Communication ProtocolSetting / Reading the max input current .(24H/25H) Setting / Reading max input voltage(22H/23H)Selecting the input on/off state (21H) Setting / Reading max input power(26H/27H) Setting / Reading CV mode voltage value.(2CH/2DH) Setting / Reading CC mode current value(2AH/2BH)Setting / Reading CW mode watt value(2EH/2FH) Setting / Reading CR mode resistance value(30H/31H) 26th byte Sum code Setting /Reading the list repeat mode CH/3DH Setting / Reading the number of list steps EH/3FHSetting / Reading List file name 48H/49H Setting / Reading timer value of for Load on 50H/51H) Save / Get list file in appointed area. CH/4DHEnable/Disable Local control mode H Setting communication address 54HSelecting / Reading trigger source H/59H Sending a trigger signal to trigging the electronic load AH Reading input voltage, current, power and relative state FHEnter the calibration mode(60H) BITGetting the calibration mode state(61H) Calibrate voltage value(62H)Calibrate current value(64H) Setting / Reading calibration information(67H/68H)Store the calibration data to EEPROM(66H) For example: Restore the factory default calibration data(69H)Reading information in bar code 6BH Setting information of bar code(6CH) Service Information Limited One-Year WarrantyDeclaration of CE Conformity Savi Ranch Parkway Yorba Linda, CA