B&K 8510 Transition Testing Operation, Set set up a constant voltage from 1.5V to Max voltage

Page 28

Step 1

Press R-set

RESISTANCE =0000R

Step 2

enter a new value by using numeric keys or Rotary

RESISTANCE=*****R

 

knob to adjust the resistance value

 

Step 3

Press Enter to confirm.

0.000W R:0000R

V-set (set up a constant voltage from 1.5V to Max voltage)

8500A electronic load can be setup for a constant voltage. Constant voltage setup procedure is as following:

 

Procedure

 

Operation details

 

VFD display

 

Step 1

 

Press

V-set

 

VOLTAGE=1.500V

 

 

 

 

 

 

 

 

Step 2

 

enter a new value by using numeric keys or Rotary

 

VOLTAGE=*.***V

 

 

knob to adjust the resistance value

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Step 3

 

Press Enter

to confirm.

 

0.000W

V:3.000V

Shift + Store

Procedure

Operation details

VFD display

 

 

 

 

 

Step 1

Press

Shift

and Store

STORE 1

Step 2

Press

Enter

to confirm.

Store the relative data

Shift + Recall

Procedure

Operation details

VFD display

 

 

 

 

 

Step 1

Press

Shift

and Recall

RECALL 1

Step 2

Press

Enter

to confirm

Recall the saving data

 

 

 

 

 

On/Off input setting

Use On/Off to change the state of electronic load. Switch on to off state by press On/Off .

Transition Testing Operation

Users could switch between the two different current and voltage level settings in the transition mode; it could test the transition specialty of power supply. Users could use front panel or communication interfaceTRAN ON AND TRAN OFFto make it work or not, please setting parameters before transition operation. Include Transition setting value, Constant pulse width setting and Transition Pulse width setting and Transition testing mode. The mix pulse width is 500uS.The Max pulse width is 6S. Transition Operation only could work in CC and CV mode.

Users can choose one of the three operation modes: Continuous, Pulse and Toggling mode.

28

Image 28
Contents Instruction Manual Programmable DC Electronic Load Table of Contents Warranty Information Safety Symbols Quick Reference Front PanelRear Panel Front Panel DisplayImmediate Action Keys Front Panel MenusKey Pad TranPOWER-ON Recall Menu OperationADC Update Rate High Lowdefault Options and Accessories General InformationDocument Orientation Getting Started MapFeatures and Capabilities AccessoriesDescription Front Panel ControlsConstant Resistance CR Mode Constant Voltage CV ModeRemote Programming Operating ModesTransient Operation Constant Power CW ModeContinuous List Operation Trigger ModePulse Input Control Triggered OperationShort On/Off Protection Features Input On/OffOperation Range Over VoltageOver Power Over TemperaturePin trigger and remote sensing connectors Remote Sense Function and Trigger FunctionSaving and Recalling Settings Wiring diagram for remote senseBattery Testing OperationVon/Voff Operation Action for set Von/Voff valueInspection InstallationLocation Power Cord Outline DiagramBench Operation Input ConnectionsCheckout Procedure Turn-On CheckoutIntroduction Trigger and Remote Sensing ConnectionsFront Panel Operation Case of TroubleDisplay Front Panel DisplayPower Set set up a constant current from 0 to Max currentExample Function keysSet set up a constant power from 0 to Max power Set set up a constant resistance from 0.1Ω to 4000ΩPOWER=*.***W 000VTransition Testing Operation Set set up a constant voltage from 1.5V to Max voltageOn/Off input setting Shift + StoreContinuous Transient Operation Transition Parameter SettingPulse Transient Operation Toggled Transient Operation StepsList Operation Edit the Testing File Automatic Testing Function of Electronic LoadQuick Recall Function Enter Automatic Testing Setting the Resolution and number of decimal pointsSpecifications Range Accuracy ResolutionRemote Operation Mode Communication cableIT-E131 RS232 Communication cable IT-E132 USB Communication cableFrame Format Communication between electronic load and PCIT-E132 communication 2AH Selecting the Remote control mode(20H) Communication ProtocolSetting / Reading the max input current .(24H/25H) Setting / Reading max input voltage(22H/23H)Selecting the input on/off state (21H) Setting / Reading max input power(26H/27H) Setting / Reading CV mode voltage value.(2CH/2DH) Setting / Reading CC mode current value(2AH/2BH)Setting / Reading CW mode watt value(2EH/2FH) Setting / Reading CR mode resistance value(30H/31H) 26th byte Sum code Setting /Reading the list repeat mode CH/3DH Setting / Reading the number of list steps EH/3FHSetting / Reading List file name 48H/49H Setting / Reading timer value of for Load on 50H/51H) Save / Get list file in appointed area. CH/4DHEnable/Disable Local control mode H Setting communication address 54HSelecting / Reading trigger source H/59H Sending a trigger signal to trigging the electronic load AH Reading input voltage, current, power and relative state FHEnter the calibration mode(60H) BITGetting the calibration mode state(61H) Calibrate voltage value(62H)Calibrate current value(64H) Setting / Reading calibration information(67H/68H)Store the calibration data to EEPROM(66H) For example: Restore the factory default calibration data(69H)Reading information in bar code 6BH Setting information of bar code(6CH) Service Information Limited One-Year WarrantyDeclaration of CE Conformity Savi Ranch Parkway Yorba Linda, CA