B&K 8510 instruction manual List Operation

Page 31

3. Press

Shift

+ Trigger

switch to the current value of 10A.

4. Press

Shift

+ Trigger

, switch between 5A and 10A.

5. Press

Shift

+ Tran

again to stop the transient operation.

List Operation

Users can use the front panel keypad or the included software (PV 8500) to program the list sequence. Please refer to the software user’s guide.

The following example will help you how to do the list operation in front panel.

Trigger

 

 

 

 

0

1

2

3

4

5

 

 

List count=1

List count=2

 

 

 

 

List sequence

Steps

1)Press the button of On/Off , execute the input of Load I in OFF state.

2)

Press

Shift

+ Menu , move cursor to the option of menu of CONFIG, Press Enter into the next step

 

menu, move cursor to TRIGGER SOURCE. Press Enter and move cursor to IMMEDIATE <DEF>, setting

 

trigger source mode is panel IMMEDIATE mode.

3)

Press

Enter

to confirm.

4)Press ESC to the previous menu, move cursor to LIST SET.

5)

Press

Enter

into the next step menu .move cursor to EDIT LIST FILE.

6)

Press

Enter

into the next step menu, move cursor to CURRENT LIST, select CURRENT MODE.

7)

Press

Enter

, move cursor to REPEAT. Setup LIST is in cycle mode.

8)

Press

Enter

to confirm, setup the list steps = 5.

Enter

9)

Press

to confirm, setup step 1 current =3A.

Enter

10)

Press

to confirm, setup step 1 width=6mS.

 

11)

Repeat 7) and 8) operation , set current and width of one step 0A, 5mS;2A, 4mS;6A,2mS;0A,5mS.

12)

Press

Enter

to confirm, Menu STORE LIST FILE 1. save file in group1.

 

13)

Move cursor to Mode Set , press

Enter

to enter into the next step menu to set mode is <LIST MODE >

 

14)

Press

Enter

to confirm

 

 

 

 

 

 

 

15)

Press

ESC ,Press On/Off

,Make Load in ON state.

16)

Press

Shift

+ Trigger

, make list operation run or stop.

17)

Stop the list operation mode. Move cursor to LIST SET, press Enter , move cursor to Mode Set in option

 

menu, press

Enter , enter into next step menu. Select mode is <FIXED MODE>.

31

Image 31
Contents Instruction Manual Programmable DC Electronic Load Table of Contents Warranty Information Safety Symbols Front Panel Quick ReferenceFront Panel Display Rear PanelTran Immediate Action KeysFront Panel Menus Key PadMenu Operation POWER-ON RecallADC Update Rate High Lowdefault Getting Started Map Options and AccessoriesGeneral Information Document OrientationFront Panel Controls Features and CapabilitiesAccessories DescriptionOperating Modes Constant Resistance CR ModeConstant Voltage CV Mode Remote ProgrammingTransient Operation Constant Power CW ModeContinuous List Operation Trigger ModePulse Input Control Triggered OperationShort On/Off Over Voltage Protection FeaturesInput On/Off Operation RangeRemote Sense Function and Trigger Function Over PowerOver Temperature Pin trigger and remote sensing connectorsWiring diagram for remote sense Saving and Recalling SettingsOperation Battery TestingAction for set Von/Voff value Von/Voff OperationInspection InstallationLocation Input Connections Power CordOutline Diagram Bench OperationTrigger and Remote Sensing Connections Checkout ProcedureTurn-On Checkout IntroductionFront Panel Display Front Panel OperationCase of Trouble DisplayFunction keys PowerSet set up a constant current from 0 to Max current Example000V Set set up a constant power from 0 to Max powerSet set up a constant resistance from 0.1Ω to 4000Ω POWER=*.***WShift + Store Transition Testing OperationSet set up a constant voltage from 1.5V to Max voltage On/Off input settingContinuous Transient Operation Transition Parameter SettingPulse Transient Operation Steps Toggled Transient OperationList Operation Edit the Testing File Automatic Testing Function of Electronic LoadQuick Recall Function Enter Setting the Resolution and number of decimal points Automatic TestingRange Accuracy Resolution SpecificationsIT-E132 USB Communication cable Remote Operation ModeCommunication cable IT-E131 RS232 Communication cableFrame Format Communication between electronic load and PCIT-E132 communication 2AH Communication Protocol Selecting the Remote control mode(20H)Setting / Reading the max input current .(24H/25H) Setting / Reading max input voltage(22H/23H)Selecting the input on/off state (21H) Setting / Reading max input power(26H/27H) Setting / Reading CV mode voltage value.(2CH/2DH) Setting / Reading CC mode current value(2AH/2BH)Setting / Reading CW mode watt value(2EH/2FH) Setting / Reading CR mode resistance value(30H/31H) 26th byte Sum code Setting / Reading the number of list steps EH/3FH Setting /Reading the list repeat mode CH/3DHSetting / Reading List file name 48H/49H Save / Get list file in appointed area. CH/4DH Setting / Reading timer value of for Load on 50H/51H)Enable/Disable Local control mode H Setting communication address 54HSelecting / Reading trigger source H/59H Reading input voltage, current, power and relative state FH Sending a trigger signal to trigging the electronic load AHBIT Enter the calibration mode(60H)Calibrate voltage value(62H) Getting the calibration mode state(61H)Calibrate current value(64H) Setting / Reading calibration information(67H/68H)Store the calibration data to EEPROM(66H) For example: Restore the factory default calibration data(69H)Reading information in bar code 6BH Setting information of bar code(6CH) Limited One-Year Warranty Service InformationDeclaration of CE Conformity Savi Ranch Parkway Yorba Linda, CA