B&K 8510 instruction manual Setting / Reading List file name 48H/49H

Page 46

From 10th to 11th byte

Time value of current step (Lower bytes are in the front location, higher bytes are in the

 

later location) (1 represent 0.1mS)

 

 

From 12th to 25th byte

System reserve

26th byte

Sum code

19. Setting / Reading one of the step’s voltage and time values. (42H/43H)

1st byte

Start bit ( AAH )

2nd byte

Address (0—0XFE)

3rd byte

Command42H/43H

 

 

 

From 4th to 5th byte

Appointed one step

From 6th byte to 9th byte

Voltage value of current step (Lower bytes are in the front location, higher bytes are in the

 

later location)

 

 

From 10th to 11th byte

Time value of current step (Lower bytes are in the front location, higher bytes are in the

 

later location) (1 represent 0.1mS)

 

 

From 12th to 25th byte

System reserve

26th byte

Sum code

20. Setting / Reading one of the step’s power and time values. (44H/45H)

1st byte

Start bit ( AAH )

2nd byte

Address (0—0XFE)

3rd byte

Command44H/45H

 

 

 

4th byte

Appointed one step

From 5th to 8th byte

Power value of current step (Lower bytes are in the front location, higher bytes are in the

 

later location)

 

 

From 9th to 10th byte

Time value of current step (Lower bytes are in the front location, higher bytes are in the

 

later location) (1 represent 0.1mS)

 

 

From 11th to 25th byte

System reserve

26th byte

Sum code

21. Setting / Reading one of the step’s power and time values. (46H/47H)

1st byte

Start bit ( AAH )

2nd byte

Address (0—0XFE)

3rd byte

Command 46H/47H

 

 

 

From 4th to 5th byte

Appointed one step

From 6th to 9th byte

Resistance value of current step (Lower bytes are in the front location, higher bytes are in

 

the later location)

 

 

From 10th to 11th byte

Time value of current step (Lower bytes are in the front location, higher bytes are in the

 

later location) (1 represent 0.1mS)

 

 

From 12th to 25th byte

System reserve

26th byte

Sum code

22. Setting / Reading List file name (48H/49H)

46

Image 46
Contents Instruction Manual Programmable DC Electronic Load Table of Contents Warranty Information Safety Symbols Quick Reference Front PanelRear Panel Front Panel DisplayKey Pad Immediate Action KeysFront Panel Menus TranPOWER-ON Recall Menu OperationADC Update Rate High Lowdefault Document Orientation Options and AccessoriesGeneral Information Getting Started MapDescription Features and CapabilitiesAccessories Front Panel ControlsRemote Programming Constant Resistance CR ModeConstant Voltage CV Mode Operating ModesTransient Operation Constant Power CW ModeContinuous List Operation Trigger ModePulse Input Control Triggered OperationShort On/Off Operation Range Protection FeaturesInput On/Off Over VoltagePin trigger and remote sensing connectors Over PowerOver Temperature Remote Sense Function and Trigger FunctionSaving and Recalling Settings Wiring diagram for remote senseBattery Testing OperationVon/Voff Operation Action for set Von/Voff valueInspection InstallationLocation Bench Operation Power CordOutline Diagram Input ConnectionsIntroduction Checkout ProcedureTurn-On Checkout Trigger and Remote Sensing ConnectionsDisplay Front Panel OperationCase of Trouble Front Panel DisplayExample PowerSet set up a constant current from 0 to Max current Function keysPOWER=*.***W Set set up a constant power from 0 to Max powerSet set up a constant resistance from 0.1Ω to 4000Ω 000VOn/Off input setting Transition Testing OperationSet set up a constant voltage from 1.5V to Max voltage Shift + StoreContinuous Transient Operation Transition Parameter SettingPulse Transient Operation Toggled Transient Operation StepsList Operation Edit the Testing File Automatic Testing Function of Electronic LoadQuick Recall Function Enter Automatic Testing Setting the Resolution and number of decimal pointsSpecifications Range Accuracy ResolutionIT-E131 RS232 Communication cable Remote Operation ModeCommunication cable IT-E132 USB Communication cableFrame Format Communication between electronic load and PCIT-E132 communication 2AH Selecting the Remote control mode(20H) Communication ProtocolSetting / Reading the max input current .(24H/25H) Setting / Reading max input voltage(22H/23H)Selecting the input on/off state (21H) Setting / Reading max input power(26H/27H) Setting / Reading CV mode voltage value.(2CH/2DH) Setting / Reading CC mode current value(2AH/2BH)Setting / Reading CW mode watt value(2EH/2FH) Setting / Reading CR mode resistance value(30H/31H) 26th byte Sum code Setting /Reading the list repeat mode CH/3DH Setting / Reading the number of list steps EH/3FHSetting / Reading List file name 48H/49H Setting / Reading timer value of for Load on 50H/51H) Save / Get list file in appointed area. CH/4DHEnable/Disable Local control mode H Setting communication address 54HSelecting / Reading trigger source H/59H Sending a trigger signal to trigging the electronic load AH Reading input voltage, current, power and relative state FHEnter the calibration mode(60H) BITGetting the calibration mode state(61H) Calibrate voltage value(62H)Calibrate current value(64H) Setting / Reading calibration information(67H/68H)Store the calibration data to EEPROM(66H) For example: Restore the factory default calibration data(69H)Reading information in bar code 6BH Setting information of bar code(6CH) Service Information Limited One-Year WarrantyDeclaration of CE Conformity Savi Ranch Parkway Yorba Linda, CA