B&K 8510 Communication between electronic load and PC, Frame Format, IT-E132 communication

Page 37

The DB9 interface connector on the rear panel of electronic load is TTL voltage level; you can use the communication cable (IT-E132) to connect the DB9 interface connector of the electronic load and the USB interface connector of computer for the communication.

 

IT-E132 communication

 

PC

cable

Load

 

COMPUTER

IT IT-E131 ISOLATED COMMUNICATION CABLE

RS232 ISOLATION TTL(5V)

859666668889942311

RX

TX

INSTRUMENT

Note: Do not connect the DB9 connector of the electronic load directly to a PC or other RS232 port.

2 Communication between electronic load and PC

Before using the remote operation mode, please make sure that the baud rate and communication address in electronic load are the same as in the computer software, otherwise, the communication will fail, you can change the baud rate and communication address from the front panel or from computer.

1Address: the range is from 0 to 254,default setting is 0

2Baud rate: 4800,9600,19200 and 38400 are selectable, default setting is 4800

3Data bit8 bit

4Stop bit1

5ParityNone

Parity=None

Start Bit

8 Data Bits

Stop Bit

Stop Bit

 

3. Frame Format

Frame length is 26 bytes. Details as following:

AAH

Address

Command 4—25bytes are information content

Parity code

Description

1Start bit is AAHoccupies one byte.

2Address range from 0 to FE, occupies one byte.

3Each command occupies one byte. Following is the command details.

20H

Selecting the Remote control mode

 

 

21H

Selecting the input on/off state

 

 

22H

Setting the max input voltage

 

 

23H

Reading the max setup input voltage.

 

 

24H

Setting max input current

 

 

37

Image 37
Contents Instruction Manual Programmable DC Electronic Load Table of Contents Warranty Information Safety Symbols Front Panel Quick ReferenceFront Panel Display Rear PanelFront Panel Menus Immediate Action KeysKey Pad TranMenu Operation POWER-ON RecallADC Update Rate High Lowdefault General Information Options and AccessoriesDocument Orientation Getting Started MapAccessories Features and CapabilitiesDescription Front Panel ControlsConstant Voltage CV Mode Constant Resistance CR ModeRemote Programming Operating ModesTransient Operation Constant Power CW ModeContinuous List Operation Trigger ModePulse Input Control Triggered OperationShort On/Off Input On/Off Protection FeaturesOperation Range Over VoltageOver Temperature Over PowerPin trigger and remote sensing connectors Remote Sense Function and Trigger FunctionWiring diagram for remote sense Saving and Recalling SettingsOperation Battery TestingAction for set Von/Voff value Von/Voff OperationInspection InstallationLocation Outline Diagram Power CordBench Operation Input ConnectionsTurn-On Checkout Checkout ProcedureIntroduction Trigger and Remote Sensing ConnectionsCase of Trouble Front Panel OperationDisplay Front Panel DisplaySet set up a constant current from 0 to Max current PowerExample Function keysSet set up a constant resistance from 0.1Ω to 4000Ω Set set up a constant power from 0 to Max powerPOWER=*.***W 000VSet set up a constant voltage from 1.5V to Max voltage Transition Testing OperationOn/Off input setting Shift + StoreContinuous Transient Operation Transition Parameter SettingPulse Transient Operation Steps Toggled Transient OperationList Operation Edit the Testing File Automatic Testing Function of Electronic LoadQuick Recall Function Enter Setting the Resolution and number of decimal points Automatic TestingRange Accuracy Resolution SpecificationsCommunication cable Remote Operation ModeIT-E131 RS232 Communication cable IT-E132 USB Communication cableFrame Format Communication between electronic load and PCIT-E132 communication 2AH Communication Protocol Selecting the Remote control mode(20H)Setting / Reading the max input current .(24H/25H) Setting / Reading max input voltage(22H/23H)Selecting the input on/off state (21H) Setting / Reading max input power(26H/27H) Setting / Reading CV mode voltage value.(2CH/2DH) Setting / Reading CC mode current value(2AH/2BH)Setting / Reading CW mode watt value(2EH/2FH) Setting / Reading CR mode resistance value(30H/31H) 26th byte Sum code Setting / Reading the number of list steps EH/3FH Setting /Reading the list repeat mode CH/3DHSetting / Reading List file name 48H/49H Save / Get list file in appointed area. CH/4DH Setting / Reading timer value of for Load on 50H/51H)Enable/Disable Local control mode H Setting communication address 54HSelecting / Reading trigger source H/59H Reading input voltage, current, power and relative state FH Sending a trigger signal to trigging the electronic load AHBIT Enter the calibration mode(60H)Calibrate voltage value(62H) Getting the calibration mode state(61H)Calibrate current value(64H) Setting / Reading calibration information(67H/68H)Store the calibration data to EEPROM(66H) For example: Restore the factory default calibration data(69H)Reading information in bar code 6BH Setting information of bar code(6CH) Limited One-Year Warranty Service InformationDeclaration of CE Conformity Savi Ranch Parkway Yorba Linda, CA