CR

The selected input channel is in the

Rmt

Indicates that the electronic load is in remote

 

resistance (CR) mode.

 

state (RS-232). In the remote state, only the

 

 

 

active key is the Local key.

 

 

 

 

Tran

The selected input channel is enabled

Shift

Indicates that the shift key has been pressed.

 

for transient operation.

 

 

 

 

 

 

List

A list is initiated or running.

Lock

keyboard is the mode for password

 

 

 

 

Function keys

Main Function:

Set up a constant current input.

Set up a constant power input.

Set up a constant resistance input

Select CC, CR and CV and CW modes.

Select Current, Resistance and Voltage levels.

Set Trigger and Transient levels.

Set up front panel measurements.

Recall the stored data

Battery testing

Short-circuit testing

Entry Keys

Entry keys let you:

Enter programming values.

Increasing or decreasing setup values.

Press or select the front panel menu parameters.

Power

The Power switch turns the electronic load power on or off.

Example

I-set (set up a constant current from 0 to Max current )

Set up a constant DC current input is the first main function of programmable DC electronic load,

8500 serial electronic load provides two methods to set up the constant DC current input by using the number keyboard and the rotary button. Please see the following operation procedure.

Procedure

Operation details

VFD display

STEP 1

Press I-set

CURRENT=0.000A

 

 

STEP 2

Enter the password or jump the step 4 if

PASSWORD:

 

your password for reentering

 

 

 

 

26

Image 26
Contents Instruction Manual Programmable DC Electronic Load Table of Contents Warranty Information Safety Symbols Quick Reference Front PanelRear Panel Front Panel DisplayKey Pad Immediate Action KeysFront Panel Menus TranPOWER-ON Recall Menu OperationADC Update Rate High Lowdefault Document Orientation Options and AccessoriesGeneral Information Getting Started MapDescription Features and CapabilitiesAccessories Front Panel ControlsRemote Programming Constant Resistance CR ModeConstant Voltage CV Mode Operating ModesContinuous Constant Power CW ModeTransient Operation Pulse Trigger ModeList Operation Short On/Off Triggered OperationInput Control Operation Range Protection FeaturesInput On/Off Over VoltagePin trigger and remote sensing connectors Over PowerOver Temperature Remote Sense Function and Trigger FunctionSaving and Recalling Settings Wiring diagram for remote senseBattery Testing OperationVon/Voff Operation Action for set Von/Voff valueLocation InstallationInspection Bench Operation Power CordOutline Diagram Input ConnectionsIntroduction Checkout ProcedureTurn-On Checkout Trigger and Remote Sensing ConnectionsDisplay Front Panel OperationCase of Trouble Front Panel DisplayExample PowerSet set up a constant current from 0 to Max current Function keysPOWER=*.***W Set set up a constant power from 0 to Max powerSet set up a constant resistance from 0.1Ω to 4000Ω 000VOn/Off input setting Transition Testing OperationSet set up a constant voltage from 1.5V to Max voltage Shift + StorePulse Transient Operation Transition Parameter SettingContinuous Transient Operation Toggled Transient Operation StepsList Operation Quick Recall Function Automatic Testing Function of Electronic LoadEdit the Testing File Enter Automatic Testing Setting the Resolution and number of decimal pointsSpecifications Range Accuracy ResolutionIT-E131 RS232 Communication cable Remote Operation ModeCommunication cable IT-E132 USB Communication cableIT-E132 communication Communication between electronic load and PCFrame Format 2AH Selecting the Remote control mode(20H) Communication ProtocolSelecting the input on/off state (21H) Setting / Reading max input voltage(22H/23H)Setting / Reading the max input current .(24H/25H) Setting / Reading max input power(26H/27H) Setting / Reading CW mode watt value(2EH/2FH) Setting / Reading CC mode current value(2AH/2BH)Setting / Reading CV mode voltage value.(2CH/2DH) Setting / Reading CR mode resistance value(30H/31H) 26th byte Sum code Setting /Reading the list repeat mode CH/3DH Setting / Reading the number of list steps EH/3FHSetting / Reading List file name 48H/49H Setting / Reading timer value of for Load on 50H/51H) Save / Get list file in appointed area. CH/4DHSelecting / Reading trigger source H/59H Setting communication address 54HEnable/Disable Local control mode H Sending a trigger signal to trigging the electronic load AH Reading input voltage, current, power and relative state FHEnter the calibration mode(60H) BITGetting the calibration mode state(61H) Calibrate voltage value(62H)Store the calibration data to EEPROM(66H) Setting / Reading calibration information(67H/68H)Calibrate current value(64H) Reading information in bar code 6BH Restore the factory default calibration data(69H)For example: Setting information of bar code(6CH) Service Information Limited One-Year WarrantyDeclaration of CE Conformity Savi Ranch Parkway Yorba Linda, CA