For example: When load receiving one trigger signal, it will switch to 10A current value, and taking 10mS to return the current value of 5A.
10A |
|
|
5A | TWD | TWD |
| 10ms | 10ms |
| TRIG | TRIG |
| Pulsed Transient Operation |
Action
1. | Select the operation mode (CC,CV,CR or CP), and set a proper value, press | On/Off to turn off the load input. | ||||
2. Press | Shift | + | , set LEVER A=5A, LEVER B=10A, WIDTH B=10mS. Transition mode now is | |||
CONTINOUS. |
|
|
|
| ||
3. | Press ▼ till >PULSE. |
|
| |||
4. | Press | Enter | , set transient mode is PULSE. |
| ||
5. | Press | Shift | + | Tran | to activate the transient mode. |
|
6. | Press | Shift | + Trigger | to start another pulse. Press Shift + Trigger | , get more pulse. | |
7. | Press | Shift | + | Tran | again to stop the transient operation. |
|
8. Press On/Off , the input of load is powered on.
.
Toggled Transient Operation
In this mode, after transition operation start, Load could change the input between the main level and the transient level when toggled transient operation is in effect.
For example:
When Load receives one trigger signal, Load current will switch between 5A and 10A.
10A |
|
5A |
|
TRG | TRG |
Toggled Transient Operation |
Steps
1. Press | Shift | + |
2. Press | Shift | + Tran to activate the transient mode. |
30