B&K 8510 instruction manual Saving and Recalling Settings, Wiring diagram for remote sense

Page 19

TRIG: A TTL-compatible input that responds to external edge trigger signal. A trigger applied to this input can be used to change settings (voltage, current, resistance, etc.), toggle between settings in transient-toggle mode, or generate a pulse in transient-pulse mode.

You must set the remote sense mode in the menu before using the remote sense function.

Action:

1Press

Shift

+ Menu key into the menu

2VFD displays >CONFIG, press Enter key to confirm

3

Press

key to choose>REMOTE SENSE, press Enter key to confirm

4

Press

/

to choose>ON, press Enter key to confirm. And the remote sense function has been set.

Wiring diagram for remote sense

Input port in front panel

+ -

Sense port in rear panel

+ -

+-

UUT

Saving and Recalling Settings

The electronic load has internal registers in which settings (mode, current, voltage, resistance, transient level, etc.) are stored. Users could use Shift + Store and Shift + Recall to save and recall the relative data as following:

CC value /CW value /CR value /CV value

Transition current A value /Transition current B value /Transition voltage A value /Transition voltage B value / Transition power A value /Transition power B value /Transition Resistance A value /Transition Resistance B value Current A pulse width time/ Current B pulse width time/Voltage A pulse width time/ Voltage B pulse width time /Power A pulse width time/Power B pulse width time /Resistance A pulse width time /Resistance B width time

Transition current testing mode/Transition voltage testing mode/Transition power testing mode /Transition resistance testing mode

Max current value / Max voltage value / Max power value

Example:

1)

Press

I-set

, set a value of current or voltage, press Enter

to confirm.

2)

Press

Shift

+

Store

set a store code at random, press

Enter

to confirm.

3)

Press

Shift

+

Recall

, set the store code that you set before, press

Enter to confirm, then you can get

the number stored.

19

Image 19
Contents Instruction Manual Programmable DC Electronic Load Table of Contents Warranty Information Safety Symbols Front Panel Quick ReferenceFront Panel Display Rear PanelTran Immediate Action KeysFront Panel Menus Key PadMenu Operation POWER-ON RecallADC Update Rate High Lowdefault Getting Started Map Options and AccessoriesGeneral Information Document OrientationFront Panel Controls Features and CapabilitiesAccessories DescriptionOperating Modes Constant Resistance CR ModeConstant Voltage CV Mode Remote ProgrammingTransient Operation Constant Power CW ModeContinuous List Operation Trigger ModePulse Input Control Triggered OperationShort On/Off Over Voltage Protection FeaturesInput On/Off Operation RangeRemote Sense Function and Trigger Function Over PowerOver Temperature Pin trigger and remote sensing connectorsWiring diagram for remote sense Saving and Recalling SettingsOperation Battery TestingAction for set Von/Voff value Von/Voff OperationInspection InstallationLocation Input Connections Power CordOutline Diagram Bench OperationTrigger and Remote Sensing Connections Checkout ProcedureTurn-On Checkout IntroductionFront Panel Display Front Panel OperationCase of Trouble DisplayFunction keys PowerSet set up a constant current from 0 to Max current Example000V Set set up a constant power from 0 to Max powerSet set up a constant resistance from 0.1Ω to 4000Ω POWER=*.***WShift + Store Transition Testing OperationSet set up a constant voltage from 1.5V to Max voltage On/Off input settingContinuous Transient Operation Transition Parameter SettingPulse Transient Operation Steps Toggled Transient OperationList Operation Edit the Testing File Automatic Testing Function of Electronic LoadQuick Recall Function Enter Setting the Resolution and number of decimal points Automatic TestingRange Accuracy Resolution SpecificationsIT-E132 USB Communication cable Remote Operation ModeCommunication cable IT-E131 RS232 Communication cableFrame Format Communication between electronic load and PCIT-E132 communication 2AH Communication Protocol Selecting the Remote control mode(20H)Setting / Reading the max input current .(24H/25H) Setting / Reading max input voltage(22H/23H)Selecting the input on/off state (21H) Setting / Reading max input power(26H/27H) Setting / Reading CV mode voltage value.(2CH/2DH) Setting / Reading CC mode current value(2AH/2BH)Setting / Reading CW mode watt value(2EH/2FH) Setting / Reading CR mode resistance value(30H/31H) 26th byte Sum code Setting / Reading the number of list steps EH/3FH Setting /Reading the list repeat mode CH/3DHSetting / Reading List file name 48H/49H Save / Get list file in appointed area. CH/4DH Setting / Reading timer value of for Load on 50H/51H)Enable/Disable Local control mode H Setting communication address 54HSelecting / Reading trigger source H/59H Reading input voltage, current, power and relative state FH Sending a trigger signal to trigging the electronic load AHBIT Enter the calibration mode(60H)Calibrate voltage value(62H) Getting the calibration mode state(61H)Calibrate current value(64H) Setting / Reading calibration information(67H/68H)Store the calibration data to EEPROM(66H) For example: Restore the factory default calibration data(69H)Reading information in bar code 6BH Setting information of bar code(6CH) Limited One-Year Warranty Service InformationDeclaration of CE Conformity Savi Ranch Parkway Yorba Linda, CA