B&K 8510 instruction manual Von/Voff Operation, Action for set Von/Voff value

Page 21
to confirm
to confirm

VBattery Voltage

Min voltage

t

ILoad Sink Current

t

Von/Voff Operation

You can set voltage value Von/Voff to control the input state on/off for electronic load. When the input voltage reaches the Von value, the load’s input state is on. When the input voltage reaches the Voff value, the load’s input state is off.

Action for set Von/Voff value:

1Press Shift + Menu into menu

2VFD displays >CONFIG, presskey to choose >SYSTEM SET, press Enter to confirm

3VFD displays >MAX CURRENT SET, presskey to choose>VOLTAGE ON SET, press Enter key to

confirm

4VFD displays >VOLT.ON=0.00V, press numeric keys to set Von value (0.1V to max voltage value), press Enter

5Presskey to choose >VOLTAGE OFF SET, press Enter to confirm

6VFD displays >VOLT.OFF=0.00V, press numeric keys to set Voff value (0V to max voltage value), press Enter

21

Image 21
Contents Instruction Manual Programmable DC Electronic Load Table of Contents Warranty Information Safety Symbols Front Panel Quick ReferenceFront Panel Display Rear PanelFront Panel Menus Immediate Action KeysKey Pad TranMenu Operation POWER-ON RecallADC Update Rate High Lowdefault General Information Options and AccessoriesDocument Orientation Getting Started MapAccessories Features and CapabilitiesDescription Front Panel ControlsConstant Voltage CV Mode Constant Resistance CR ModeRemote Programming Operating ModesConstant Power CW Mode Transient OperationContinuous Trigger Mode List OperationPulse Triggered Operation Input ControlShort On/Off Input On/Off Protection FeaturesOperation Range Over VoltageOver Temperature Over PowerPin trigger and remote sensing connectors Remote Sense Function and Trigger FunctionWiring diagram for remote sense Saving and Recalling SettingsOperation Battery TestingAction for set Von/Voff value Von/Voff OperationInstallation InspectionLocation Outline Diagram Power CordBench Operation Input ConnectionsTurn-On Checkout Checkout ProcedureIntroduction Trigger and Remote Sensing ConnectionsCase of Trouble Front Panel OperationDisplay Front Panel DisplaySet set up a constant current from 0 to Max current PowerExample Function keysSet set up a constant resistance from 0.1Ω to 4000Ω Set set up a constant power from 0 to Max powerPOWER=*.***W 000VSet set up a constant voltage from 1.5V to Max voltage Transition Testing OperationOn/Off input setting Shift + StoreTransition Parameter Setting Continuous Transient OperationPulse Transient Operation Steps Toggled Transient OperationList Operation Automatic Testing Function of Electronic Load Edit the Testing FileQuick Recall Function Enter Setting the Resolution and number of decimal points Automatic TestingRange Accuracy Resolution SpecificationsCommunication cable Remote Operation ModeIT-E131 RS232 Communication cable IT-E132 USB Communication cableCommunication between electronic load and PC Frame FormatIT-E132 communication 2AH Communication Protocol Selecting the Remote control mode(20H)Setting / Reading max input voltage(22H/23H) Setting / Reading the max input current .(24H/25H)Selecting the input on/off state (21H) Setting / Reading max input power(26H/27H) Setting / Reading CC mode current value(2AH/2BH) Setting / Reading CV mode voltage value.(2CH/2DH)Setting / Reading CW mode watt value(2EH/2FH) Setting / Reading CR mode resistance value(30H/31H) 26th byte Sum code Setting / Reading the number of list steps EH/3FH Setting /Reading the list repeat mode CH/3DHSetting / Reading List file name 48H/49H Save / Get list file in appointed area. CH/4DH Setting / Reading timer value of for Load on 50H/51H)Setting communication address 54H Enable/Disable Local control mode HSelecting / Reading trigger source H/59H Reading input voltage, current, power and relative state FH Sending a trigger signal to trigging the electronic load AHBIT Enter the calibration mode(60H)Calibrate voltage value(62H) Getting the calibration mode state(61H)Setting / Reading calibration information(67H/68H) Calibrate current value(64H)Store the calibration data to EEPROM(66H) Restore the factory default calibration data(69H) For example:Reading information in bar code 6BH Setting information of bar code(6CH) Limited One-Year Warranty Service InformationDeclaration of CE Conformity Savi Ranch Parkway Yorba Linda, CA