B&K 8510 instruction manual Quick Recall Function, Automatic Testing Function of Electronic Load

Page 32

18) Press

Enter

to confirm.

 

Quick Recall Function

This function allows you to recall 10 groups numbers which were stored previously.

Steps:

1) Press Shift + Menu

2)

Press▼to CONFIG, press Enter into submenu, press▼ to SHORTCURT RECALL

3)

Press Enter , select ON and press Enter to confirm.

4)

Press ESC

two times and quit the menu.

5)

Press 1

and you can get the first group number. Press 2 , you can get the second group number. Press 0 ,

you can get the tenth group number. And if there is not stored number, VFD will show NO EEPROM DATA.

Automatic Testing Function of Electronic Load

The CD load’s firmware provides an automatic testing function that is convenient for testing various parameters of power supplies with different load conditions. Now we take a kind of charger as an example to explain the usage of the automatic testing of the electronic load.

The test proceeding of R35 charger

Work produce

process

 

 

 

 

method

 

 

 

 

 

 

 

 

 

 

Mode

Voltage(V)

Current(mA)

Power(W)

Ripple

 

 

 

 

 

 

 

counter(mVpp)

 

 

 

 

 

 

 

 

The first step

CC

5.86.15

210

≤4(input175VAC)

 

 

 

 

 

 

 

 

 

 

The second step

None

5.96.4

 

0

≤1.2(input175VAC)

 

 

 

 

 

 

 

 

 

The third step

CC

5

205245

 

 

 

 

 

 

 

 

 

≤50mVpp

 

The fourth step

CV

3

205

 

 

245

 

 

 

The fifth step

CV

2

205245

 

 

 

 

 

 

 

 

 

 

The sixth step

Short

0

≤245

 

 

 

 

circuit

 

 

 

 

 

 

 

 

 

 

 

 

 

Edit the Testing File

1) Press Shift + Menu into menu, VFD displays CONFIG

2)Press▼ to LIST SET.

3) Press Enter into submenu, press▼ toEDIT LIST FILE.

32

Image 32
Contents Instruction Manual Programmable DC Electronic Load Table of Contents Warranty Information Safety Symbols Quick Reference Front PanelRear Panel Front Panel DisplayImmediate Action Keys Front Panel MenusKey Pad TranPOWER-ON Recall Menu OperationADC Update Rate High Lowdefault Options and Accessories General InformationDocument Orientation Getting Started MapFeatures and Capabilities AccessoriesDescription Front Panel ControlsConstant Resistance CR Mode Constant Voltage CV ModeRemote Programming Operating ModesContinuous Constant Power CW ModeTransient Operation Pulse Trigger ModeList Operation Short On/Off Triggered OperationInput Control Protection Features Input On/OffOperation Range Over VoltageOver Power Over TemperaturePin trigger and remote sensing connectors Remote Sense Function and Trigger FunctionSaving and Recalling Settings Wiring diagram for remote senseBattery Testing OperationVon/Voff Operation Action for set Von/Voff valueLocation InstallationInspection Power Cord Outline DiagramBench Operation Input ConnectionsCheckout Procedure Turn-On CheckoutIntroduction Trigger and Remote Sensing ConnectionsFront Panel Operation Case of TroubleDisplay Front Panel DisplayPower Set set up a constant current from 0 to Max currentExample Function keysSet set up a constant power from 0 to Max power Set set up a constant resistance from 0.1Ω to 4000ΩPOWER=*.***W 000VTransition Testing Operation Set set up a constant voltage from 1.5V to Max voltageOn/Off input setting Shift + StorePulse Transient Operation Transition Parameter SettingContinuous Transient Operation Toggled Transient Operation StepsList Operation Quick Recall Function Automatic Testing Function of Electronic LoadEdit the Testing File Enter Automatic Testing Setting the Resolution and number of decimal pointsSpecifications Range Accuracy ResolutionRemote Operation Mode Communication cableIT-E131 RS232 Communication cable IT-E132 USB Communication cableIT-E132 communication Communication between electronic load and PCFrame Format 2AH Selecting the Remote control mode(20H) Communication ProtocolSelecting the input on/off state (21H) Setting / Reading max input voltage(22H/23H)Setting / Reading the max input current .(24H/25H) Setting / Reading max input power(26H/27H) Setting / Reading CW mode watt value(2EH/2FH) Setting / Reading CC mode current value(2AH/2BH)Setting / Reading CV mode voltage value.(2CH/2DH) Setting / Reading CR mode resistance value(30H/31H) 26th byte Sum code Setting /Reading the list repeat mode CH/3DH Setting / Reading the number of list steps EH/3FHSetting / Reading List file name 48H/49H Setting / Reading timer value of for Load on 50H/51H) Save / Get list file in appointed area. CH/4DHSelecting / Reading trigger source H/59H Setting communication address 54HEnable/Disable Local control mode H Sending a trigger signal to trigging the electronic load AH Reading input voltage, current, power and relative state FHEnter the calibration mode(60H) BITGetting the calibration mode state(61H) Calibrate voltage value(62H)Store the calibration data to EEPROM(66H) Setting / Reading calibration information(67H/68H)Calibrate current value(64H) Reading information in bar code 6BH Restore the factory default calibration data(69H)For example: Setting information of bar code(6CH) Service Information Limited One-Year WarrantyDeclaration of CE Conformity Savi Ranch Parkway Yorba Linda, CA