18) Press | Enter | to confirm. |
|
Quick Recall Function
This function allows you to recall 10 groups numbers which were stored previously.
Steps:
1) Press Shift + Menu
2) | Press▼to CONFIG, press Enter into submenu, press▼ to SHORTCURT RECALL | |
3) | Press Enter , select >ON and press Enter to confirm. | |
4) | Press ESC | two times and quit the menu. |
5) | Press 1 | and you can get the first group number. Press 2 , you can get the second group number. Press 0 , |
you can get the tenth group number. And if there is not stored number, VFD will show NO EEPROM DATA.
Automatic Testing Function of Electronic Load
The CD load’s firmware provides an automatic testing function that is convenient for testing various parameters of power supplies with different load conditions. Now we take a kind of charger as an example to explain the usage of the automatic testing of the electronic load.
The test proceeding of R35 charger
Work produce | process |
|
|
|
| method |
|
|
|
|
|
|
|
| |
|
| Mode | Voltage(V) | Current(mA) | Power(W) | Ripple | |
|
|
|
|
|
|
| counter(mVpp) |
|
|
|
|
|
|
| |
| The first step | CC | 5.8~6.15 | 210 | ≤4(input175VAC) |
| |
|
|
|
|
|
|
|
|
| The second step | None | 5.9~6.4 |
| 0 | ≤1.2(input175VAC) |
|
|
|
|
|
|
|
| |
| The third step | CC | 5 | 205~245 |
|
| |
|
|
|
|
|
|
| ≤50mVpp |
| The fourth step | CV | 3 | 205 | ~ |
| |
| 245 |
|
| ||||
| The fifth step | CV | 2 | 205~245 |
|
| |
|
|
|
|
|
|
| |
| The sixth step | Short | 0 | ≤245 |
|
| |
|
| circuit |
|
|
|
|
|
|
|
|
|
|
|
|
|
Edit the Testing File
1) Press Shift + Menu into menu, VFD displays >CONFIG |
2)Press▼ to >LIST SET.
3) Press Enter into submenu, press▼ to>EDIT LIST FILE.
32