B&K 8510 instruction manual Setting /Reading the list repeat mode CH/3DH

Page 45

From 10th byte to 13th byte

Setting value of resistance B (Lower bytes are in the front location, higher bytes are in

 

the later location)

 

 

From 14th byte to 15th byte

Time value of timer B (Lower bytes are in the front location, higher bytes are in the later

 

location) (1 represent 0.1mS)

 

 

16th byte

Transition operation mode (0 is CONTINUES,1 is PULSE,2 is TOGGLED)

17th byte to 25th byte

System reserve

26th byte

Sum code

15Selecting /Reading the list operation mode (CC/CV/CW/CR)3AH/3BH

1st

byte

Start bit ( AAH )

2nd

byte

Address (0—0XFE)

3rd

byte

Command3AH/3BH

 

 

 

 

4th byte

LIST operation mode0is CC mode1 is CV mode ,2 is CW mode,3 is CR mode

 

 

 

 

From 5th to 25 byte

System reserve

26th byte

Sum code

16. Setting /Reading the list repeat mode. (3CH/3DH)

1st byte

Start bit ( AAH )

2nd byte

Address (0—0XFE)

3rd byte

Command3CH/3DH

 

 

 

4th byte

LIST repeat operation mode(0 is ONCE, 1 is REPEAT)

From 5th to 25th byte

System reserve

26th byte

Sum code

17. Setting / Reading the number of list steps. (3EH/3FH)

1st byte

Start bit ( AAH )

2nd byte

Address (0—0XFE)

3rd byte

Command3EH/3FH

 

 

 

From 4thto 5th byte

LIST steps

From 6th to 25th byte

System reserve

26th byte

Sum code

18. Setting / Reading one of the step’s current and time values. (40H/41H)

1st byte

Start bit ( AAH )

2nd byte

Address (0—0XFE)

3rd byte

Command40H/41H

 

 

 

From 4th byte to 5th byte

Appointed one step

From 6th to 9th byte

Current value of current step (Lower bytes are in the front location, higher bytes are in the

 

later location)

 

 

45

Image 45
Contents Instruction Manual Programmable DC Electronic Load Table of Contents Warranty Information Safety Symbols Front Panel Quick ReferenceFront Panel Display Rear PanelFront Panel Menus Immediate Action KeysKey Pad TranMenu Operation POWER-ON RecallADC Update Rate High Lowdefault General Information Options and AccessoriesDocument Orientation Getting Started MapAccessories Features and CapabilitiesDescription Front Panel ControlsConstant Voltage CV Mode Constant Resistance CR ModeRemote Programming Operating ModesConstant Power CW Mode Transient OperationContinuous Trigger Mode List OperationPulse Triggered Operation Input ControlShort On/Off Input On/Off Protection FeaturesOperation Range Over VoltageOver Temperature Over PowerPin trigger and remote sensing connectors Remote Sense Function and Trigger FunctionWiring diagram for remote sense Saving and Recalling SettingsOperation Battery TestingAction for set Von/Voff value Von/Voff OperationInstallation InspectionLocation Outline Diagram Power CordBench Operation Input ConnectionsTurn-On Checkout Checkout ProcedureIntroduction Trigger and Remote Sensing ConnectionsCase of Trouble Front Panel OperationDisplay Front Panel DisplaySet set up a constant current from 0 to Max current PowerExample Function keysSet set up a constant resistance from 0.1Ω to 4000Ω Set set up a constant power from 0 to Max powerPOWER=*.***W 000VSet set up a constant voltage from 1.5V to Max voltage Transition Testing OperationOn/Off input setting Shift + StoreTransition Parameter Setting Continuous Transient OperationPulse Transient Operation Steps Toggled Transient OperationList Operation Automatic Testing Function of Electronic Load Edit the Testing FileQuick Recall Function Enter Setting the Resolution and number of decimal points Automatic TestingRange Accuracy Resolution SpecificationsCommunication cable Remote Operation ModeIT-E131 RS232 Communication cable IT-E132 USB Communication cableCommunication between electronic load and PC Frame FormatIT-E132 communication 2AH Communication Protocol Selecting the Remote control mode(20H)Setting / Reading max input voltage(22H/23H) Setting / Reading the max input current .(24H/25H)Selecting the input on/off state (21H) Setting / Reading max input power(26H/27H) Setting / Reading CC mode current value(2AH/2BH) Setting / Reading CV mode voltage value.(2CH/2DH)Setting / Reading CW mode watt value(2EH/2FH) Setting / Reading CR mode resistance value(30H/31H) 26th byte Sum code Setting / Reading the number of list steps EH/3FH Setting /Reading the list repeat mode CH/3DHSetting / Reading List file name 48H/49H Save / Get list file in appointed area. CH/4DH Setting / Reading timer value of for Load on 50H/51H)Setting communication address 54H Enable/Disable Local control mode HSelecting / Reading trigger source H/59H Reading input voltage, current, power and relative state FH Sending a trigger signal to trigging the electronic load AHBIT Enter the calibration mode(60H)Calibrate voltage value(62H) Getting the calibration mode state(61H)Setting / Reading calibration information(67H/68H) Calibrate current value(64H)Store the calibration data to EEPROM(66H) Restore the factory default calibration data(69H) For example:Reading information in bar code 6BH Setting information of bar code(6CH) Limited One-Year Warranty Service InformationDeclaration of CE Conformity Savi Ranch Parkway Yorba Linda, CA