5 Key Reference

data evaluated during that time and present a single point of trace data based on the detector mode. We call the interval during which data for that trace point is being collected, the “bucket”. Thus a trace is more than a series of single points. It is actually a series of trace “buckets”. The data may be sampled many times within each bucket.

With the detector sets to Auto, Pressing [TRACE] > {More (1 of 2)} > {Average} and select On changes the

detector. The Auto choice depends on marker, and trace functions, and the trace averaging function. If a marker function or measurement is running, the Auto choice of detector is either Average or Sample. When one of the detectors (such as Average) is manually selected instead of Auto, that detector is used without regard to other analyzer settings.

Auto

Selects peak detector as the default. If a measurement condition arises where a different type of detector scheme would be better utilized, the system will use the alternate scheme.

Key access: [TRACE] > {More (1 of 2)} > {Detector}

• Normal

Selects the maximum and minimum video signal values alternately. When selecting the normal detector, “N” appears in the upper- left corner.

Key access: [TRACE] > {More (1 of 2)} > {Detector}

• Average

Measures the average value of the amplitude across each trace interval (bucket). When Average is

selected, “A” appears on the left corner.

Key access: [TRACE] > {More (1 of 2)} > {Detector}

92

N9340A User’s Guide

Page 102