4 System Setting

Perform a Time Base Calibration

 

As soon as the calibration function triggers, it

 

interrupts the current measurement is interrupted

 

and a gauge displays on the LCD. The gauge simply

 

indicates calibration action rather than calibration

 

course, as the calibration time is unpredictable.

 

When the calibration is finished, the LCD displays

 

a calibration, and the interrupted measurement

 

restarts.

 

Time base calibration takes during a short time only when the

N O T E

inner temperature is stable. When the inner temperature is

 

 

increasing, calibration takes a long-time course or will fail. If

 

the input reference signal is abnormal, the calibration cycle

 

will take a long and unpredictable time to exit, and the LCD

 

displays an error message.

 

 

 

Key Access: [SYS] > {More (2 of 3)} > {Calibration}