Key Reference 5
Limit
Limit Line
Activates an adjustable horizontal line that is used as a visual reference line.
The line, which can be used for trace arithmetic, has amplitude values that correspond to its vertical position when compared to the reference level. The value of the display line appears in the active function block and on the left side of the display. Use the arrow keys, knob, or numeric keypad to adjust the display line. Pressing any digit, 0 through 9, on the numeric keypad brings up the
selected terminator menu. To deactivate the display line, press Display Line (Off).
Key access: [Limit]
Limit Pattern
Turns on the limit pattern function and turns off the limit line function simultaneously.
Key access: [Limit]
Set Pattern
Allows you to set a limit pattern for measurements.
You can create or edit limit points up to four points to define the limit line.
Pressing X Axis Unite to toggle the frequency domain or time domain before you set the parameter of limit point.
Key access: [Limit]
Limit Type
Defines the limit you are editing as either an upper or lower limit. An upper limit fails if trace 1 exceeds the limit. A lower limit fails if trace 1 falls below the limit.
Key access: [Limit]
N9340A User’s Guide | 95 |