Tracking Generator Subsystem

The tracking generator subsystem controls the use tracking generator function of the analyzer.

State

:TGENerator:AMPLitude:STATe OFFON01 :TGENerator:AMPLitude:STATe?

This command toggles the state of tracking generator between on and off. Setting to ON state enables the tracking generator mode.

*RST State:

OFF

Key Access:

1

Press [MODE] to enable the tracking

 

 

generator mode.

 

2

Press {Amplitude} to toggle the tracking

 

 

generator between on and off.

 

 

 

Output Amplitude

:TGENerator:AMPLitude <ampt> :TGENerator:AMPLitude?

This command sets the tracking generator output amplitude. This command is valid when tracking generator is enabled.

*RST State:

–20 dBm

Range:

0 to –30 dBm

Available unit:

dBm

Key Access:

{Amplitude}

 

 