CALCulate:MARKer Subsection

SCPI Command Reference 6

CALCulate:MARKer Subsection

Markers All Off on All Traces

:CALCulate:MARKer:AOFF

This command turns off all markers on all traces.

Key access:

[Marker] > {More} > {All Off}

Markers All On

:CALCulate:MARKer:ALL

This command turns on all the markers and places them on six highest peaks.

*RST state:

Off

Key access:

[Marker] > {Peak search} > {Auto search}

 

 

Frequency Counter Marker

:CALCulate:MARKer[1]23456:FCOunt [:STATe] OFFON01

:CALCulate:MARKer[1]23456:FCOunt

[:STATe]?

This command toggles the frequency counter between on and off.

*RST state:

Off

Key access:

[Marker] > {More} > {Freq Count}

 

 

To query frequency counter readout, use

:CALCulate:MARKer[1]23456:FRECount:X?